Kwaliteit van Medische Laboratoria
GLOBAAL RAPPORT EKE HbA1c VERS BLOED - 2016.1A
Target HbA1c = 62 mmol/mol (Referentiemethode NGSP/DCCT)
INSTRUMENT
RESULT
(Mean)
N
SD
CV%
Bias
Bias %
Capillarys 2 flex – piercing IFCC mmol/mol 59 11 1.7 2.9 -3.0 -4.9
D-10_A1c_Program IFCC mmol/mol 63 6 2.2 3.4 1.1 1.8
D-100_A1c_Program IFCC mmol/mol 62 2 0.7 1.1 0.0 0.0
Menarini HA-8140 DM IFCC mmol/mol 65 1 ****
*
***** 3.5 5.6
Menarini HA-8160 TP IFCC mmol/mol 63 7 1.8 2.8 1.3 2.2
Menarini HA-8160 VP IFCC mmol/mol 62 34 4.7 7.6 0.3 0.4
Menarini HA-8180V IFCC mmol/mol 63 29 1.3 2.0 1.0 1.6
Minicap flex-piercing mmol/mol 59 6 0.8 1.4 -3.0 -4.9
Tina-quant, Whole blood, Gen.2, IFCC
mmol/mol
63
2
0.7
1.1
1.0
1.6
Tina-quant, Whole blood, Gen.3, IFCC
mmol/mol
63
4
1.7
2.8
1.0
1.6
TOSOH G7 IFCC mmol/mol 61 2 1.4 2.3 -0.5 -0.9
TOSOH G8 IFCC mmol/mol 63 42 1.3 2.1 1.4 2.2
Variant II_HbA1c IFCC mmol/mol 61 2 0.0 0.0 -0.5 -0.9
Variant-II_Turbo_A1c_Program IFCC
mmol/mol
62
3
1.0
1.6
0.5
0.8
Opmerking:
Om redenen van correcte verwerking van de EKE-resultaten is het belangrijk dat alle laboratoria in dezelfde eenheid
(mmol/mol) rapporteren. Ter herinnering, de overschakeling van de HbA1c eenheden naar mmol/mol heeft plaats gevonden in juni 2011.