Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON...

13
Till SOHIER , Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON, Hubert GRESS, Mauricio TORRES, Quentin RAFFY Université de Strasbourg et Institut Pluridisciplinaire Hubert Curien Département de Recherches en Subatomiques Laboratoire de Chimie Nucléaire

Transcript of Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON...

Page 1: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON, Hubert GRESS, Mauricio TORRES, Quentin RAFFY

Université de Strasbourg et Institut Pluridisciplinaire Hubert Curien

Département de Recherches en Subatomiques

Laboratoire de Chimie Nucléaire

Page 2: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

• GOAL:

Improvement of the Linear Energy Transfert (LET) measurement in the hadrontherapy field

• Comparison:

Between ionisation chamber (reference method) and fluorescence intensity emission measurement

• Problem:

Direct measurement methods are not reliable because of the high noise and saturation of the photodetectors

• Solution:

Time correlated single photon counting (TCSPC) method

2

Summary

Page 3: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Experimental methods: Time correlated single photon counting (TCSPC) method

Impulsion de

synchronisation

Détection

d’un photon

t = t1-t0

temps

Résultat après

plusieurs évenements

Forme typique

d’un déclin

3

PM2

PM1

E

PM: photomultiplicateur CTA: convertisseur temps-amplitude AMC: analyseur multicanal

excitation

ionisation

Logarith

mic

scale

Time (nanosecond)

Fluorescence decay

Page 4: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Fluorescent Samples

4

Polyéthylène téréphtalate (PET) [Goodfellow©]

Polyvinyltoluene (PVT) [Bicron©]

Page 5: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

• Experiments were lead at the Heavy Ion Medical Accelerator in Chiba (HIMAC) of the National Institute of Radiobiological Sciences (NIRS).

Fluorescence measurement for several LET : Protocols

5 5

Page 6: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Fluorescence measurement for several LET : Protocols

6

Page 7: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Fluorescence measurement for several LET : Protocols

7

Sensitivity

• Reference method: gives the result directly in Grays • Can be considered as an absolut measurement method

Page 8: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Fluorescence intensity measurement for several LET : Results

8

• Normalized to irradiation time

Fluorescence decays at different LET

Page 9: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Fluorescence measurement for several LET : Results

9

Irradiation under carbon beam of 290 MeV/u

PMMA Thickness (mm)

LET (

arb

itra

ry u

nits)

PET under Carbon beam

Page 10: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Fluorescence measurement for several LET : Results

10

Irradiation under hélium beam of 150 MeV/u

LET (

arb

itra

ry u

nits)

PMMA Thickness (mm)

PVT under Helium beam

Page 11: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Results summary

• TCSPC measurement method is more sensitive than the ionisation chamber for carbon and helium

• TCSPC method seems to show higher LET sensitivity than predicted by the Bethe-Block formula with carbons

11

Measurement method

Direct measure

Ionisation Chamber

SRIM TCSPC

Sensitivity carbon

2,44 4,91 7,27 11,4

Sensitivity helium

3,79 6,38 4,52

Page 12: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Discussion

12

I(t)

t

facteur 10

facteur 1000

à 10 000

• TCSPC avoids saturation through statiscal reconstruction of the fluorescence decay

Page 13: Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON ...iramis.cea.fr/radiolyse/Colloque2011/talks/sohier.pdf · Till SOHIER, Mélodie MUNIER, Jean-Marc JUNG, Rémi BARILLON,

Conclusion

• TCSPC measurement methods avoids the intrinsic Photomultiplier noise

Noise is not correlated in coïncidence method

• better sensivity in comparison to the traditional method (ionisation chamber)

Because Fluorescence takes also in account the excitation in the medium

• This method can also be used to measure, at high resolution, the dose rate of X and gamma rays.

13