Optimization and development of a second-dimension ... · poriën van de deeltjes van de kolom,...
Transcript of Optimization and development of a second-dimension ... · poriën van de deeltjes van de kolom,...
Bachelor Thesis Scheikunde
Optimization and development of a second-dimension separation
method for the analysis of polymer nanoparticles by
comprehensive two-dimensional liquid chromatography
Door
Ruben Kers
1 juli 2016
Studentnummer
10499776
Onderzoeksinstituut Verantwoordelijke docent
Van ’t Hoff Institute for Molecular Sciences Prof. dr. ir. P.J. Schoenmakers
Onderzoeksgroep Begeleider
Analytical Chemistry Group B.W.J. Pirok, MSc.
2
Table of contents
List of Abbreviations .............................................................................................................................. 4
Abstract / Samenvatting .......................................................................................................................... 5
1. Introduction ......................................................................................................................................... 7
2. Theory ................................................................................................................................................. 9
2.1 Liquid chromatography (LC) ........................................................................................................ 9
2.1.1 The principle of LC ............................................................................................................... 9
2.1.2 Resolution and band broadening ......................................................................................... 10
2.2 Size-exclusion chromatography (SEC) ....................................................................................... 11
2.2.1 The principle of SEC ........................................................................................................... 11
2.2.2 Practical aspects of SEC ...................................................................................................... 12
2.2.3 Core-shell columns .............................................................................................................. 15
2.3 Hydrodynamic Chromatography (HDC) .................................................................................... 15
2.3.1 Principle of HDC ................................................................................................................. 15
2.3.2 Practical aspects of HDC ..................................................................................................... 17
3. Experimental Section ........................................................................................................................ 18
3.1 Chemicals ................................................................................................................................... 18
3.2 Instrumental ................................................................................................................................ 18
3.3 Analytical Methods ..................................................................................................................... 19
3.3.1 Sample Preparation ............................................................................................................. 19
3.3.2 Separation of PLGA with different SEC columns .............................................................. 21
3.3.3 Calibration of size-exclusion columns ................................................................................ 21
3.3.4 Analysis of large polydispersity PS ..................................................................................... 22
3.3.4 HDC column testing ............................................................................................................ 22
4. Results & Discussion ........................................................................................................................ 23
4.1 Separation of PLGA with PLGel columns ................................................................................. 23
4.1.1 Column efficiency determination with PLGA .................................................................... 23
4.1.2 SEC calibration with polystyrene and polymethylmethacrylate ......................................... 28
4.2 Separation of PLGA mixture with experimental core-shell columns ......................................... 32
4.2.1 Column efficiency determination with PLGA .................................................................... 32
4.2.2 SEC calibration with polymethylmethacrylate .................................................................... 34
4.2.3 SEC Calibration with 3 coupled columns ........................................................................... 35
4.2.4 Analysis of large polydispersity PS ..................................................................................... 36
4.2.5 Concluding remarks regarding SEC analysis ...................................................................... 37
4.3 HDC Measurements ................................................................................................................... 38
5. Conclusion ........................................................................................................................................ 41
6. Future Prospects ................................................................................................................................ 42
References ............................................................................................................................................. 43
3
Acknowledgement
First of all, I would like to thank Bob Pirok, MSc. for his guidance and supervision during my
bachelor project, his insight and willingness to share his views with me on the different
problems were most supportive and certainly increased the quality of my work. Secondly, I
want to thank Fleur van Beek, MSc. for being accessible and helpful with my project whenever
I was in need of it. I would also like to thank prof. dr. ir. Peter Schoenmakers and dr. Wim Kok
for allowing me to do my project at the Analytical Chemistry Group at the UvA. Finally, I
would like to thank Serafine, Pascal, Charlotte and Leon for helping me with the practical
aspects of HPLC.
4
List of Abbreviations
3-NBS acid 3-Nitrobenzenesulfonic acid
BHT Butylated hydroxytoluene
Dc Column diameter
DAD Diode array detector
e Porosity
γ Unadjusted relative retention
GFC Gel filtration chromatography
GPC Gel permeation chromatography
H Plate height
HDC Hydrodynamic chromatography
HPLC High-performance liquid chromatography
IEC Ion-exchange chromatography
L Column length
LC Liquid chromatography
LC×LC Two-dimensional liquid chromatography
N Plate number
NPLC Normal-phase liquid chromatography
PLGA Poly(lactic-co-glycolic acid)
PMMA Polymethylmethacrylate
PS Polystyrene
RID Refractive-index detector
RPLC Reversed-phase liquid chromatography
SC Slalom chromatography
SEC Size-exclusion chromatography
tR Retention time
THF Tetrahydrofuran
Ux Linear flowrate
UHPLC Ultra-high performance liquid chromatography
V0 Void volume
VR Retention volume
5
Abstract
The separation of molecules by size-exclusion chromatography (SEC) is nowadays a widely
established technique in the determination of molecular weight distribution of polymers, but
combining SEC with hydrodynamic chromatography (HDC) in one comprehensive analytical
two-dimensional LC system (LC×LC) for the determination of polymer nanoparticles has not
yet been achieved. In this study, which is part of the MAnIAC project, research is conducted
towards the development of a second dimension method for this comprehensive analytical
system, which is based on SEC and where HDC is applied to verify the proposed method. From
this study it can be concluded that columns with high plate numbers have to be used in order
to achieve good resolution when separating low-molecular weight polymers. It has been found
that using multiple columns and certain core-shell columns result in sufficient separation
efficiency to give good resolution between the investigated samples. HDC analysis on
polymers has not been performed as method verification due to insufficient column
performance.
Samenvatting
Vloeistofchromatografie is een scheidingsmethode die binnen de analytische chemie veel
wordt toegepast voor de analyse van chemische stofmengsels. Bij vloeistofchromatografie
wordt een scheiding uitgevoerd in een kolom, gevuld met deeltjes, waar doorheen een continue
stroom van vloeistof loopt. Daarin wordt het te analyseren monster opgelost. Deze oplossing
komt uiteindelijk in een detector terecht, waarbij - in het ideale geval - de individuele stoffen
kunnen worden onderscheiden. Een scheidingskolom bestaat meestal uit deeltjes van vaste stof,
zoals silica of koolstof. Bepaalde stoffen hebben een hogere affiniteit of passen beter door de
poriën van de deeltjes van de kolom, waardoor er een scheiding wordt gedaan op basis van
bijvoorbeeld hydrofobiteit, lading of grootte van moleculen.
In de analyse en scheiding van polymeren voor de bepaling van hun molmassa, wordt
tegenwoordig vaak gebruik gemaakt van methoden als size exclusionchromatografie (SEC) en
hydrodynamische chromatografie (HDC). Het is echter nog moeilijk om beide
scheidingsmethoden in één systeem aan elkaar te koppelen. In theorie zou daardoor meer
informatie verkregen kunnen worden over de te analyseren stof dan wanneer slechts één
methode wordt toegepast. In het MAnIAC project wordt getracht een systeem te ontwerpen dat
dit soort analyses uit kan voeren en wel met behulp van HDC en SEC.
6
In dit bachelor project ligt de focus op het vinden van een goede chromatografische methode
voor het scheiden van polymeren met behulp van SEC. Hierbij wordt vooral gekeken naar welk
type kolommen het beste werken voor het scheiden van polymeren met verschillende
deeltjesgrootte. Het is gebleken dat een goede scheiding plaatsvindt tussen polymeren met
relatief lage molmassa wanneer meerdere scheidingskolommen worden toegepast of
kolommen worden gebruikt die deeltjes bevatten met een harde kern en poreuze laag. Ook is
geprobeerd om HDC toe te passen voor de verificatie van de SEC-methode. Uit de resultaten
is gebleken dat de HDC-kolom in onvoldoende mate functioneerde om goede analyse uit te
voeren. Voor vervolgonderzoek zou dus de methode geverifieerd kunnen worden met HDC als
scheidingsmethode.
Chapter 1: Introduction
7
1. Introduction
Polymers are large molecules that play key roles in everyday life and are interesting due to
their wide range of properties. Polymers consist of an array of repeating subunits of carbon
backbone which can contain various functional groups and heteroatoms. Examples of widely
known polymers are DNA, as a naturally occurring polymer and polystyrene as a synthetically
occurring polymer. Polymers are used in a broad variety of applications due to their variable
chain size, hydrophobicity and functions, and are therefore incorporated in a large array of
different materials.1
An important application of polymers is the synthesis of nanoparticles, which are
particles with at least one dimension between 1 nm and 100 nm in size.2 An example of an
application of nanoparticles is in the use of paint as coatings that have anti-reflective
properties.3 These coatings are furthermore applied in for instance anti-corrosion solutions,
enhanced surface appearance for furniture and UV-protection.4 For companies that produce
coatings and the polymer nanoparticles they consist of, it is important to know the composition
of their product and its characteristics. The knowledge on how their product behaves can be
used to provide an opportunity for further development of the product. Conventionally used
analytical techniques are applied to analyze these nanoparticles and polymers, such as
chromatography.5
In the MAnIAC project, research is conducted towards the development of a
comprehensive analytical system to determine the most important characteristics of these
nanoparticles. The proposed system is portrayed in Figure 1, and is based on the principle of
two-dimensional liquid chromatography (LC×LC) where two distinct separations are
combined in one system for improved chromatographic results.
Figure 1: Proposed comprehensive analytical system under development in the MAnIAC
project.
Chapter 1: Introduction
8
In the first dimension it is envisaged to separate the polymeric nanoparticles based on their size
by using hydrodynamic chromatography (HDC). The nanoparticles are then to be dissolved in
the modulator and the products of the dissolution are analyzed by a second dimension
separation such as size-exclusion chromatography (SEC), reversed-phase liquid
chromatography (RPLC) or ion-exchange chromatography (IEC).
HDC and SEC/RPLC are the methods of choice in this comprehensive system due to
the important information that is acquired with these techniques. HDC is capable of separating
aggregated nanoparticles in aqueous solution, while SEC is highly efficient in determining the
polymers in organic solution that make up the larger nanoparticles. The proposed MAnIAC
system could therefore achieve a higher selectivity compared to currently used methods, and is
thus exceptionally interesting for certain applications in industry such as polymer
characterization.
In this study, a possible second dimension separation mechanism is investigated on
latex nanoparticles and polymers, using SEC and HDC. The results from this study can be used
to fine-tune the system operating conditions like used stationary phases, flow rates, injection
volumes, and dilution factors for the second dimension separation. In this research, the
measurements for chromatographic optimization will be conducted using polystyrene (PS),
polymethylmethacrylate (PMMA) and poly(lactic-co-glycolic acid) (PLGA) samples with
differing chain lengths to measure the response of the analytical system.
Nanoparticles will be prepared with building blocks of PLGA, by making different
known ratios of building blocks (consisting of PLGA and polyethylene glycol (PEG)) and
separating them in the first dimension of the comprehensive system with HDC in aqueous
mobile phase. The nanoparticles are then to be dissolved and separated in the second dimension
with SEC in THF mobile phase, where the original building blocks can be distinguished and
by which the method is verified.
It is expected that the separation of these polymers is challenging due to the wide
molecular weight distribution of the analyzed molecules, resulting in broad peaks and
consequently loss of resolution.
Chapter 2: Theory
9
2. Theory
2.1 Liquid chromatography (LC)
2.1.1 The principle of LC
Liquid chromatography (LC) is a widely used separation technique which separates analytes
based on their characteristics, such as hydrophobicity or molecular size. A typical LC setup
consists of a pump, column and detector. In LC, the separation occurs in the column (the
stationary phase) which consists of either a nonpolar material like C18, or a polar material like
silica for respectively reversed-phase liquid chromatography (RPLC) or normal-phase liquid
chromatography. Another mode of separation in LC is based on the molecular size of the
analytes, which is the separation mechanism that occurs in size-exclusion chromatography
(SEC) and hydrodynamic chromatography (HDC). The analyte is dissolved in the mobile
phase, which usually consists of either acetonitrile, methanol, water, THF or mixtures of
solvent that flows continuously through the system.6 In Figure 2 a typical setup of an LC
system is shown.
Figure 2: Schematic of an LC setup.7
When pressures of up to 400 bar are applied for separation in LC, the technique is called high-
performance liquid chromatography (HPLC). When operating under ultra-high pressures
(~1000 bar) the technique is referred to as ultra-high performance liquid chromatography
(UHPLC). The benefit of using high pressure is the fact that separation can be performed faster,
and smaller particles in the stationary phase can be used which increases resolution at the
expense of increased backpressure.8
Chapter 2: Theory
10
LC is nowadays used as a method of analysis or as a purification method, due to the
fractionation that occurs in the column.6 The LC output can be subsequently coupled to
different types of detection methods, like UV detection, refractive-index detection,
fluorescence detection, evaporate light scattering detection or mass spectrometry (optical
rotation detection, tandem mass spectrometry).6 In this study, HPLC is applied with UV and
refractive-index detection.
2.1.2 Resolution and band broadening
The resolution of separation in LC depends on two factors: the difference in elution time of
compounds and the band broadening of peaks. There are multiple approaches to influencing
the resolution, for example by changing the elution strength of the mobile phase, changing the
temperature of the column, changing particle size of the column and changing the column
length.6 In general, the higher the amount of theoretical plates, the better the resolution (see Eq.
(1)):
𝑅𝑒𝑠𝑜𝑙𝑢𝑡𝑖𝑜𝑛 √𝑁
𝛾 − (1)
Where N is the number of theoretical plates, and γ is the unadjusted relative retention.6
The amount of plates on a column is related to the length of the column and the plate height by
Eq. (2):
𝑁𝐿
𝐻 (2)
Where N is the number of plates, L is the column length and H is the plate height.6
The influence of the flow rate on the plate number and plate height in a certain column can
furthermore be described by the Van Deemter Equation (Eq. (3)):
𝐻 ≈ 𝐴 𝐵
𝑈𝑥 𝐶𝑈𝑥 (3)
Where H is the plate height, A, B and C are constants dependent on the multiple flow paths
within the column, longitudinal diffusion and equilibration time, respectively. Ux is the linear
velocity.6
Chapter 2: Theory
11
The Van Deemter Curve describes the column efficiency of the stationary phase and is
dependent on the band broadening processes that occur during separation. Band broadening
occurs for instance due to Eddy diffusion, which is related to the multiple flow paths that
molecules take when going through the column. Flowrates are different in narrow and broad
flow paths, which causes different velocities in the same column and results in band
broadening. The A-term in the Van Deemter curve describes this Eddy diffusion. Furthermore,
longitudinal diffusion causes band broadening due to dispersion of analyte within the column,
which is described by the B-term in the Van Deemter Curve. Also the resistance of the analyte
to mobile-phase mass transfer is an important cause of band broadening, which means that
solute near the particles of the packing move more slowly through the column than the solute
that is further away from the packing. This is described by the C-term of the Van Deemter
curve. The characteristics of the column packing, like particle size and pore sizes have
significant influence on band broadening processes. To optimize resolution, it is therefore
important to take into account the type of stationary phase used.9
The resolution of separations, plate number and plate height are important factors in the
development of useful LC×LC analysis methods. However, the magnitude of these effects are
often times best to be determined experimentally and therefore the aim of this study is to find
optimal conditions in which good polymer separations can be performed.
2.2 Size-exclusion chromatography (SEC)
2.2.1 The principle of SEC
Size exclusion chromatography (SEC) is a variant of chromatography in which compounds are
separated based on their size. The stationary phase consists of porous packing material, in
which small particles diffuse through pores of the column packing, whereas larger molecules
are excluded of the pores. There is no interaction with the stationary phase other than the
selective diffusion through the pores that is caused by entropic effects.9 The result is a
separation based on the difference in rate at which molecules migrate through pores of the
stationary phase particles, which depends on the molecular weight and hydrodynamic size of
the molecules.5, 10 This process of size exclusion during a SEC separation is illustrated in Figure
3.
Chapter 2: Theory
12
Figure 3: Illustrative schematic of exclusion of compounds within pores in SEC.
SEC is often applied for the determination of the average molecular weight of biopolymers and
synthetic polymers. There are various types of SEC columns, they differ in for example pore
size, column length and column diameter. In gel permeation chromatography (GPC), a
stationary phase consists of a hydrophobic gel in combination with an apolar mobile phase,
while in gel filtration chromatography (GFC), a hydrophilic packing material is used in
combination with a polar mobile phase. Packing material with distinct pore sizes are produced
for optimal separation of different polymer size ranges and there are multiple types of columns
available for a broad series of polymers.11
2.2.2 Practical aspects of SEC
A typical size determination measurement in SEC starts with the setup of a calibration curve
with known molecular weight standards. The logarithm of the molar weight is plotted against
the retention volume (VR), from which the molecular weight of the unknown sample is
calculated. A calibration curve is different for every system, due to the multiple factors
involved that determine retention in a column. Parameters which are of significant influence
on the measurements are for example the size of the packing material, the material of which
the packing material consists, the inner diameter and the length of the used columns.9
Furthermore, the hydrodynamic size of the analyzed polymers has to be taken into account and
the right polymer standards need to be used, as polymers have different ways in which they are
folded.5 Calibration measurements are applied in this study for the calculation of polymer
molar mass, and to confirm that the utilized SEC method provides good results with known
PLGA samples. An example of a SEC calibration curve is portrayed in Figure 4.
Chapter 2: Theory
13
Figure 4: Example of a SEC calibration curve of polystyrene.12
The maximum size of molecules that can be separated on a SEC column elute on the left side
of the calibration curve that is following an increasing vertical trend, which is termed the
exclusion limit of the column and is caused by the fact that those molecules are of sizes too
large to enter the pores of the stationary phase and thus go through the column without being
retained. The other end of the calibration curve is called the permeation limit, which is achieved
when molecules have such a low molar mass that they penetrate the pores of the stationary
phase completely. The t0 marker, which in this study is toluene in SEC, elutes at the permeation
limit.11 The V0 and therefore t0 in SEC can be predicted theoretically with Eq. (4):
𝑉 𝐷𝑐𝐿𝑒
𝜋 (4)
Where V0 is the void volume, dc is the column diameter, L is the length of the column and e is
the porosity.13
When molecules reach a certain hydrodynamic size above the exclusion limit of a
column, they can start to unfold and stretch within the stationary phase, which causes a
phenomenon that is called slalom chromatography (SC). SC is caused by the stretching of
molecules turning around the stationary phase packing material, which results in increased
retention and is an entirely different retention mechanism than that of SEC. The tendency in
which SC occurs in the stationary phase is dependent on the size of the analyzed molecules,
the column packing, the flow rate, solvent viscosity and temperature of the column.9 In this
study, SC is observed in the calibration of columns with polymer standards, but not during
PLGA analysis due to the small molecular weight of the samples used in this study.
Chapter 2: Theory
14
Current SEC methods show promising results in determining the characteristics of polymers,
and it is therefore an increasingly important technique in the determination of molar weight of
particles in the 100 Å to 10.000 Å linear dimension range.10 Furthermore, one of the features
of polymer analysis with SEC is that the molecular weight distribution can be determined, due
to the differing polydispersity of the polymers and elution with a Gaussian distribution. Both
biopolymers and synthetic polymers diverge from narrowly to broadly polydisperse, which
results in varying broadness of peaks on the chromatogram. Due to the wide distribution of the
molar weight, there is potential loss of resolution which poses a challenge in choosing the right
stationary phases in SEC.9 Figure 5 depicts a mass spectrum of polystyrene, showing the
distribution of polymer molecular weight caused by the polydispersity, resulting in broad peaks
in SEC chromatograms.
Figure 5: MALDI-ToF-MS spectrum of polystyrene with a molecular weight of 5050 g/mol and
polydispersity of 1.05.14
Chapter 2: Theory
15
2.2.3 Core-shell columns
An emerging new technology in HPLC is the application of core-shell particles in SEC
separations. Core-shell particles are particles of porous silica or C18 material which is wrapped
around a solid core, resulting in a particle size of ~2.6 μm, but which has the efficiency of a
sub-2 μm particle size column. This increased efficiency is induced by the lower pore volume
in core-shell columns, which lowers the longitudinal diffusion in the column, but also decreases
the A- and the C-term in the Van Deemter curve due to faster mass transfer and rougher surface
of the particles. The disadvantage of core-shell columns is the fact that the pore volume is
significantly smaller than for columns with completely porous particles, which retarded the use
of core-shell columns due to fear of lower separation capability. But on the contrary, the latest
studies on core-shell columns show that these stationary phases are readily capable of excellent
resolution and fast separations on UHPLC devices.
In this study, experimental Phenomenex silica and C18 core-shell columns will be
applied in the separation of polymers, to investigate if these core-shell columns are indeed
capable of increased resolution in PLGA separations.
2.3 Hydrodynamic Chromatography (HDC)
2.3.1 Principle of HDC
A chromatographic method which is exceptionally useful in the determination of nanosize
particles is HDC, in which non-porous inert particles in packed columns or open tube capillary
columns are applied to induce separation, and in which the size, the shape and the structure of
analytes can be determined, when coupled to a variety of detection methods. The separation in
HDC is based on the Poiseuille-like (parabolic) flow profile, which is the result of different
flow lines within the column. Flow lines that are closer to the walls of the column or to the
stationary phase particles move slower than flow lines in the middle of the column. Due to the
capability of smaller molecules to approach the walls or particles of the column more closely,
they elute at a later time compared to the larger particles that experience a greater net flow.10
Figure 6 on the next page depicts the mechanism of HDC separation.
Chapter 2: Theory
16
Figure 6: Mechanism of HDC separation in an open tubular column and a packed column.10
HDC is originally an early established technique in chromatography, but has undergone a
resurgence because of its capability in separating molecules with higher molar mass than
SEC.10 The use of HDC columns is attractive when large hydrophobic polymers are being
analyzed in aqueous samples, due to micelle formation which is the result of the aggregation
of hydrophobic material in order to reduce its contact area with the hydrophilic environment.
The micelles itself are larger particles than the un-aggregated polymers, scaling up their
hydrodynamic size into the nanoparticle range and thus making HDC an excellent technique
for separation.10 Figure 7 portrays a schematic of a polymer micelle.
Figure 7: Micelle which is build up from polymers in aqueous solution.16
Chapter 2: Theory
17
2.3.2 Practical aspects of HDC
In HDC, packed columns are more conventionally used due to the narrow channel that is
necessary in open HDC columns. Using open HDC columns is difficult, which is especially
caused by the requirement of low injection volumes (~1 µL) to avoid column overloading.5
HDC effects arise during separations of very large molecules and for columns with
small pore sizes, which is a widely occurring phenomenon due to the fact that at a certain point
of molecular weight the molecules approach the exclusion limit and will stop transfering
through the porous stationary phase. It is therefore possible to perform HDC with size-
exclusion columns. The rate at which HDC effects occur depends on the ratio of the pore size
of the particles to the diameter of the particles (Rp/dp), a decrease in this ratio results in less
HDC effects in the stationary phase and a more dominant SEC separation mechanism during
the separation.9 For instance in this study, HDC effects are induced with experimental SEC
columns that contain small porous particles, on molecules with high molecular weight.
In the determination of molar weight distribution of polymers with HDC, it is standard
procedure to first construct a calibration. A calibration is performed by injecting polymers of
known sizes and measuring the retention time of the polymer with corresponding molar weight,
just like in SEC. An illustration of different calibration curve regions for HDC, SEC and SC is
given below in Figure 8.17
Figure 8: Calibration with polystyrene showing the specific molar weight regions in which
they are separated using SEC, HDC and SC mechanisms.17
Chapter 3: Experimental Section
18
3. Experimental Section
3.1 Chemicals
Tetrahydrofuran (THF, AR-grade, stabilized and HPLC-S grade, unstabilized) was obtained
from Biosolve B.V. (Valkenswaard, the Netherlands) and used for the preparation of samples
and as SEC mobile phase. Butylated hydroxytoluene (BHT) was obtained from Sigma Aldrich
(St. Louis, MO, USA) and was applied as HPLC-S grade THF stabilizer. Toluene was acquired
from Biosolve B.V. (Valkenswaard, the Netherlands) and was applied as SEC marker.
Deionised water (Arium 611UV, Sartorius, R=18.2 MΩ cm-1, Germany) was prepared as HDC
mobile phase and applied in sample dilutions. HDC buffer was obtained from Agilent
Technologies (Amstelveen, the Netherlands) and used as HDC mobile phase surfactant.
Potassium dichromate was purchased from Sigma Aldrich (St. Louis, MO, USA), uracil was
obtained from Sigma Aldrich (St. Louis, MO, USA) and 3-nitrobenzenesulfonic acid (3-NBS
acid) was received from Sigma Aldrich (St. Louis, MO, USA), which were used as an HDC
marker. Polystyrene (PS) and poly(methylmethacrylate) (PMMA) samples with different
polymer length were obtained from Polymer Laboratories (Amherst, U.S.). Poly(lactic-co-
glycolic acid) (PLGA) samples with a molar weight distribution of 21 kDa and 10.5 kDa and
large polydispersity PS with molecular weight distribution of 200 kDa were received from
DSM Coating Resins (Waalwijk, the Netherlands).
3.2 Instrumental
All HPLC measurements using RID detection were conducted on a combined Agilent 1100
series and 1260 series LC setup, which consisted of an 1100 series nanopump (G2226A), an
1100 series refractive-index detector (G1362A), an Infinity 1260 series degasser (G1322A) and
an 1100 series thermostatted column compartment (G1316A). Manual injection was used with
a 20 μL sample loop installed.
All HPLC measurements using UV detection were conducted on a combined Agilent
1100 series, 1260 series and 1290 series LC setup, which consisted of a 1100 series quaternary
pump (G1311A), a 1290 series diode array detector (G4212A), an Infinity 1260 series degasser
(G1322A), and an 1100 series thermostatted column compartment (G1316A). An autosampler
was used with 20 µL injections, consisting of an 1100 series ALS (G1313A).
Chapter 3: Experimental Section
19
All UHPLC measurements were conducted on an Agilent 1290 Infinity LC system, which
consisted of an Infinity 1290 binary pump (G4220A), an Infinity 1290 diode-array detector
(G4212A), an Infinity 1290 autosampler and an Infinity 1290 thermostatted column
compartment (G1316C).
3.3 Analytical Methods
3.3.1 Sample Preparation
SEC calibration standards were prepared separately by dissolving 100 mg of either PS or
PMMA with known molecular weight in 5 mL of THF, to a concentration of 20 mg/mL. The
SEC calibration stock solutions were subsequently diluted in THF to a concentration of 0.5
mg/mL. The molecular weight of both used PS and PMMA standards are given on the next
page in Table 1.
Chapter 3: Experimental Section
20
Table 1: Molecular weight of the used PS and PMMA SEC calibration standards.
PS
Standard
Molar
Weight
(g/mol)
PMMA
Standard
Molar Weight
(g/mol)
PS01 580 PMMA01 620
PS02 980 PMMA02 1140
PS03 1990 PMMA03 1310
PS09 9920 PMMA04 1960
PS10 13880 PMMA05 2000
PS11 19880 PMMA06 2870
PS12 52400 PMMA07 6820
PS13 70950 PMMA08 10260
PS16 299400 PMMA09 14920
PS19 735500 PMMA10 24830
PS21 1373000 PMMA11 30690
PS22 2061000 PMMA12 49600
PS23 2536000 PMMA13 79500
PS24 3053000 PMMA14 100000
PS25 3743000 PMMA15 141500
PS26 7450000 PMMA16 211400
PS27 13000000 PMMA17 300300
- - PMMA18 518900
- - PMMA19 659000
- - PMMA20 948500
- - PMMA21 1250000
PLGA samples for SEC were prepared by dissolving 64.3 mg of (PLGA(7.5 kDa))2-PEG(6
kDa) (PLGA 21 kDa) in 1 mL THF, 63.4 mg (PLGA(3.75kDa))2-PEG(3 kDa) (PLGA 10.5
kDa) in 1 mL THF and dissolving 30.2 mg PLGA (21 kDa) combined with 31.1 mg PLGA
(10.5 kDa) in 1 mL THF. The samples were subsequently diluted 10, 50 and 100 times in THF
to a volume of 1 mL.
Chapter 3: Experimental Section
21
A vial of 3-Nitrobenzenesulfonic acid (3-NBS acid) was prepared to a concentration of 2.0
mg/mL in HDC buffer. 301 mg Potassium dichromate was dissolved in 5 mL HDC buffer
after which the solution was diluted to a concentration of 0.301 mg/mL in HDC buffer. 200
mg Uracil was dissolved in 5 mL warm HDC buffer and diluted to a concentration of 2.0
mg/mL. The prepared solutions were used as HDC t0 marker.
3.3.2 Separation of PLGA with different SEC columns
To determine whether the PLGA particles could efficiently be separated with the available
PLGel columns, different PLGA sample mixtures were injected into the HPLC system using
PLGel Mixed-E (7.5x100 mm, 3 μm), PLGel Mixed-D (7.5x300 mm, 5 μm) and PLGel Mixed
(7.5x600 mm) columns, using refractive-index detection. The flowrate was set to 1.0 mL/min.
The mobile phase consisted of 100% THF (AR-grade, stabilized). The column compartment
and the RID had a temperature set to 30 °C. The injection volume was 20 μL for every
measurement.
To test the efficiency of a Phenomenex core-shell B109 column (4.6x150 mm, 2.6 μm,
XB-C18), the different PLGA sample mixtures were injected into the HPLC system using
refractive-index detection. The flowrate was set to 0.2 mL/min. The mobile phase consisted of
100% THF (HPLC-S grade, stabilized). The column compartment and the RID had a
temperature set to 30 °C. The injection volume was 20 µL for every measurement.
3.3.3 Calibration of size-exclusion columns
Calibrations with PS and PMMA were performed by injecting every PS and PMMA standard
(see Table 1) into the HPLC setup with RID detection. A PMMA and PS calibration was
conducted with PLGel Mixed (7.5x600 mm), with a flowrate of 1.0 mL/min for every
measurement. The mobile phase consisted of 100% THF (AR-grade, Stabilized). The used
column and RID temperature was set to 30 °C. The injection volume was 20 µL for every
measurement.
A calibration was performed on a Phenomenex core-shell B109 column (4.6x150 mm,
2.6 μm, XB-C18), by injecting PMMA standards (see Table 1) with the HPLC setup on RID
detection. A flowrate of 0.2 mL/min was used for every measurement. The mobile phase
consisted of 100% THF (HPLC-S grade, unstabilized) with added BHT (to 200 ppm). The used
column and RID temperature was set to 30 °C. The injection volume was 20 µL for every
measurement.
Chapter 3: Experimental Section
22
A calibration with PS (see Table 1) was performed on three coupled Phenomenex core-shell
B111 (4.6x150 mm, 3.6 μm, XB-C18), B104 (4.6x150 mm, 2.6 μm, silica) and B109 (4.6x150
mm, 2.6 μm, XB-C18) columns, on an UHPLC system with UV detection. The flowrate was set
to 2.00 mL/min for every measurement. The mobile phase consisted of 100% THF (HPLC-S
grade, unstabilized) with added BHT (to 200 ppm). The column compartment had a
temperature set to 25 °C. The injection volume was 1 µL for every measurement.
3.3.4 Analysis of large polydispersity PS
To see how the Phenomenex core-shell B111, B104 and B109 columns respond to large
polydispersity polymers, a large polydispersity PS sample with a molecular weight of 200.000
g/mol was injected on an UHPLC system with UV detection. The flowrate was set to 2.00
mL/min for every measurement. The mobile phase consisted of 100% THF (HPLC-S grade,
unstabilized) with added BHT (to 200 ppm). The column compartment had a temperature set
to 25 °C. The injection volume was 1 µL for every measurement.
3.3.4 HDC column testing
An injection with 3-NBS acid and uracil was performed using an Agilent PL-PSDA Cartridge
Type 1 HDC column (7.5x800 mm), on an HPLC system with UV detection. The flowrate
was set to 1.6 mL/min for the 3-NBS acid injection, uracil injection and a blank sample,
containing acidified HDC buffer (pH ~3.2). The mobile phase consisted of acidified HDC
buffer. The used column temperature was set to 25 °C. The injection volume was 0.5 μL for
3-NBS acid and 5 μL for uracil.
An injection with potassium dichromate was performed using an Agilent PL-PSDA
Cartridge Type 1 HDC column (7.5x800 mm), on an HPLC system with UV detection. The
flowrate was set to 1.7 mL/min for both the potassium dichromate injection and the blank
sample, containing HDC buffer (pH ~7). The mobile phase consisted of HDC buffer. The
used column temperature was set to 25 °C. The injection volume was 10 μL.
Chapter 4: Results & Discussion
23
4. Results & Discussion
4.1 Separation of PLGA with PLGel columns
4.1.1 Column efficiency determination with PLGA
Figure 9: Chromatogram of PLGA mixture, blank THF sample and toluene marker, separated
with a PLGel Mixed-E column.
Figure 9 shows the chromatogram of the PLGA mixture sample separated on a PLGel Mixed-
E column. As illustrated in the figure, there is a signal visible at 1.99 min. which is assigned to
the PLGA sample, it is absent in the blank sample and elutes before the toluene marker, which
elutes at 3.09 min. The t0 marker has a retention time that sufficiently corresponds to the
theoretically predicted value of 3.27 min., which is the retention time at the permeation limit
and is calculated according to Eq. (4). The positive signal at 2.73 min. and the strong negative
signals are suspected system peaks. The positive system peaks are most likely the result of
components present in the sample but absent in the eluent, which could be due to either
contamination of the sample or the formation of peroxides in the sample that form when the
THF reacts with oxygen. The negative signals are caused by compounds in the eluent that are
not present in the injection, which could for example be peroxides that exist in a different
concentration than in the sample.18 Figure 10 depicts the chromatogram of this PLGA sample.
-8000
-6000
-4000
-2000
0
2000
4000
6000
0 1 2 3 4 5 6
Sign
al (
nR
IU)
Time (min)
PLGel Mixed-E Column
Blank
t0
PLGA
Chapter 4: Results & Discussion
24
Figure 10: Chromatogram of PLGA mixture, separated with a PLGel Mixed-E column.
There is no separation between both PLGA samples visible from this chromatogram, but a
broad distribution is observed which is assumed to correspond to both PLGA samples. From
this result it can be concluded that the used column was not capable of giving sufficient
resolution for the separation between the PLGA samples labeled to be 10.5 kDa and 21 kDa.
To improve separation using this type of column, a column with a larger length could
potentially be applied for the separation of the PLGA particles, to increase the plate number
according to Eq. (2). To confirm that the low resolution is the cause of low column efficiency
in the separation of these PLGA samples, a calibration curve could be constructed, which shows
that an insufficient resolution is achieved. This particular calibration curve of the PLGel Mixed-
E column was not made in this project.
-200
0
200
400
600
800
1,5 1,6 1,7 1,8 1,9 2 2,1 2,2 2,3 2,4 2,5
Sign
al (
nR
IU)
Time (min)
PLGel Mixed-E Column
PLGA
Chapter 4: Results & Discussion
25
Figure 11: Chromatogram of PLGA mixture, blank THF sample and toluene marker,
separated with a PLGel Mixed-D column.
Figure 11 shows the chromatogram of the same PLGA sample mixture, this time separated on
a PLGel Mixed-D column, which was tested due to its molar weight operating range of 200-
400.000 g/mol and longer column length compared to the PLGel Mixed-E column, which in
theory results in a higher plate number.19 It is assumed that the signal at 6.76 min. is
corresponding to the PLGA mixture, due to its absence in the blank and because of elution
before the toluene marker at 9.47 min. As can be observed from this figure, the PLGA sample
mixture once again shows a broad distribution which is expected due to the polydispersity of
PLGA polymers. Furthermore, it is assumed that the negative signals shown in the
chromatogram are system peaks which is most likely the result of compounds in the THF
mobile phase, which might be due to the formation of peroxides. Furthermore, from this figure
it can be seen that the toluene marker has a significantly higher intensity of negative signals
compared to the blank and PLGA samples. This can be related to the fact that the used vial
with toluene, was prepared from a different bottle of THF solvent than the blank and PLGA
sample. From this chromatogram it can be concluded that the composition of this THF was
different. In Figure 12 below, a chromatogram is given which highlights the PLGA signal.
-5000
-4000
-3000
-2000
-1000
0
1000
2000
3000
4000
0 2 4 6 8 10 12 14 16
Inte
nsi
ty (
nR
IU)
Time (min)
PLGel Mixed-D Column
Blank
t0
PLGA
Chapter 4: Results & Discussion
26
Figure 12: Chromatogram of PLGA mixture, separated with a PLGel Mixed-D column.
Using this chromatogram, it can be concluded that the PLGA mixture cannot be sufficiently
separated with this column, due to the substantial broadness of the signals, resulting in low
resolution. There is a small bump visible at 6.5 min which might correspond to the larger
molecular weight PLGA sample, but there is a substantial overlap between both compounds.
To increase separation, a column with higher column efficiency for this molecular weight range
should be used.
Figure 13: Chromatogram of PLGA mixture, blank THF sample and toluene marker,
separated with a PLGel Mixed double column.
-100
-50
0
50
100
150
200
250
300
5 5,5 6 6,5 7 7,5 8
Inte
nsi
ty (
nR
IU)
Time (min)
PLGel Mixed-D Column
PLGA
-4000
-2000
0
2000
4000
6000
8000
10000
0 5 10 15 20 25
Inte
nsi
ty (
nR
IU)
Time (min)
PLGel Mixed 2 columns
Blank
t0
PLGA
Chapter 4: Results & Discussion
27
Figure 13 shows the chromatogram of the PLGA mixture, separated on two PLGel mixed
columns. As can be observed from the chromatogram, the signal at 14.5 min corresponds to
the PLGA mixture, which is determined from the PLGA sample and the blank mixture. It is
assumed that the positive signal at 18.8 min and the negative signals are system peaks, which
might be the result of the formation of peroxides due to aging of the THF solvent. Figure 14
below highlights the PLGA signal.
Figure 14: Chromatogram of the PLGA mixture and individual injections, separated with a
PLGel Mixed Double column.
As illustrated in this chromatogram, there is a broad signal visible which is assigned to the 21
kDa PLGA sample, and a signal with less intensity at ~15.5 min. that overlaps with this broad
peak of higher intensity. It is assumed this signal corresponds to the PLGA with a mass of 10.5
kDa, due to the overlap with the individual injection. The lower intensity of this smaller PLGA
sample might be the result of either higher polydispersity, pipetting error or insufficient
homogenization of both samples. Nevertheless, these results suggest that the PLGel Mixed
double column can induce adequate separation of these PLGA particles.
-200
-100
0
100
200
300
400
12 13 14 15 16 17
Inte
nsi
ty (
nR
IU)
Time (min)
PLGel Mixed 2 columns
PLGA 10.5 kDa
PLGA 21 kDa
PLGA Mixture
Chapter 4: Results & Discussion
28
4.1.2 SEC calibration with polystyrene and polymethylmethacrylate
Figure 15: Chromatograms of PMMA with molar weight distributions of 2000 and 24830
g/mol.
Figure 15 depicts the chromatogram of PMMA polymer calibration standards with molar
weight distributions of 2000 g/mol, and 24830 g/mol used with a double column of PLGel
Mixed (7.5x600 mm), recorded with RID detection. As portrayed in the chromatogram, there
are signals at 14.3 and 16.2 min. which have been assigned to the PMMA standards. The
positive signal at 18.0 min. and the negative signals are assumed to be system peaks related to
degradation of THF to peroxides in the prepared samples and the mobile phase, and correspond
to the blank sample which consisted of THF. Furthermore, as can be seen from this
chromatogram, the signal corresponding to the smaller polymer standard elutes after the larger
polymer standard, which is expected due to the fact that smaller molecules enter the pores of
the packing more easily and are therefore more retained in SEC.
-3000
-2500
-2000
-1500
-1000
-500
0
500
1000
1500
2000
0 5 10 15 20 25
Inte
nsi
ty (
nR
IU)
Time (min)
PLGel Mixed 2 Columns
PMMA 2000 g/mol
PMMA 24830 g/mol
Chapter 4: Results & Discussion
29
Figure 16: Chromatograms of PS with molar weight distributions of 1990 and 19880 g/mol.
Figure 16 depicts the chromatogram of PS calibration standards with molar weight
distributions of 1990 and 19880 g/mol, recorded with the same setup as the PMMA calibration.
As is depicted in this figure, there are positive signals at 13.8 and 16.0 min. corresponding to
the PS samples, which is determined from the subtraction of the blank sample. The positive
signal at 17.9 min. and the negative signals are assumed to be system peaks, which occur due
to aging of THF in the prepared samples and in the mobile phase.
-8000
-6000
-4000
-2000
0
2000
4000
0 5 10 15 20 25
Inte
nsi
ty (
nR
IU)
Time (min)
PLGel Mixed 2 columns
PS 1990 g/mol
PS 19880 g/mol
Chapter 4: Results & Discussion
30
Figure 17: Calibration curves of PLGel Mixed double column, created with PMMA and PS
standards.
Figure 17 shows the calibration curves of both the PMMA calibration and the PS calibration.
As can be seen from this figure, both calibration curves show a linear trend after the permeation
limit corresponding to the SEC region of the curve, up to a molar weight of about 107 g/mol
and are quite comparable. This result was expected due to the fact that even though PMMA
and PS have different hydrodynamic sizes, the retention mechanism is for both standards the
same in SEC. From these calibration curves it can be concluded that both PS and PMMA can
be used to determine PLGA molecular weights under these conditions.
Table 2: PLGA molecular weight results calculated according to PMMA and PS calibration
curve fitting given in Figure 17.
Calibration PLGA 21 kDa PLGA 10.5 kDa
PMMA 5758.52 g/mol 11005.14 g/mol
PS 5733.71 g/mol 10751.71 g/mol
0
1
2
3
4
5
6
7
8
0 5 10 15 20
Log
MW
VR (mL)
PMMA & PS calibration curves of PLGel Mixed 2 columns
PMMA Calibration
PS Calibration
Chapter 4: Results & Discussion
31
In Table 2, the results are given of the molecular weight calculations with the injected PLGA
samples of 21 kDa and 10.5 kDa, using a third order fitting. As can be concluded from this
table, the calculations for PLGA with a molar weight of 10.5 kDa are fairly accurate, while the
calculation results for PLGA with a molar weight of 21 kDa show largely different numbers
than what was expected.
There are multiple scenarios that could explain the results, the PLGA corresponding to
the molar weight of 21 kDa could have decayed, which would have resulted in a lower
molecular mass and is a feasible explanation due to the biodegradable nature of PLGA.20 The
other scenario is that the PS and PMMA calibrations were not representative for the PLGA
molecules due to the different ways that the molecules fold, although this case is less likely
because of the fact that both calibrations give nearly the same results. Another potential issue
could be that the PLGA of 21 kDa was labeled incorrectly, or that the PLGA sample
corresponding to the 21 kDa molecular weight was deposited into the wrong vial, which could
explain the higher viscosity of the PLGA which was labeled to be the 10.5 kDa PLGA (as
higher molecular weight polymers generally have a higher viscosity than lower molecular
weight polymers).5, 21
Chapter 4: Results & Discussion
32
4.2 Separation of PLGA mixture with experimental core-shell columns
4.2.1 Column efficiency determination with PLGA
Figure 18: Chromatogram of PLGA mixture, blank THF sample and toluene marker,
separated with the Phenomex core-shell B109 column.
Figure 18 depicts the chromatogram of the 10.5 kDa and 21 kDa PLGA mixture, as well as the
toluene and blank sample. As is shown in the chromatogram, there are broad signals visible at
6.03 and 6.76 min., which are assigned to the PLGA mixture due to the absence of these signals
in the blank sample. The signal which is shown as the PLGA and blank sample at 7.14 min., as
well as the negative peaks are assumed to be system peaks, caused by the aging of THF. The
signal at 7.00 min. is assigned to the toluene marker, as it corresponds to the predicted
permeation limit calculated by Eq. (4). The signal at 7.39 min. is expected to be either an
impurity in the toluene sample or caused by aging of the sample, as the toluene sample was
prepared from a different THF bottle than the blank and PLGA samples, and can therefore
show different signals.
-60000
-40000
-20000
0
20000
40000
60000
0 2 4 6 8 10 12 14 16
Inte
nsi
ty (
nR
IU)
Time (min)
Phenomenex core-shell B109
Blank
t0
PLGA
Chapter 4: Results & Discussion
33
Figure 19: Chromatogram of the PLGA mixture and individual PLGA injections, separated
with the Phenomenex core-shell B109 column.
Figure 19 shows the chromatogram of the PLGA mixture and individual PLGA injections, with
fixed y- and x-axis at the PLGA signals. There is a good separation visible between the 10.5
kDa and 21 kDa-labeled PLGA samples at 6.03 and 6.76 min. It has been determined that these
signals indeed correspond to the used PLGA samples, by comparing them to the separate
injections of both polymers. It can be observed that baseline separation had not been realized,
but the Phenomenex core-shell B109 column achieves the highest resolution from all
investigated columns, which can be related to the fact that the stationary phase consisted of
small core-shell particles, resulting in lower chromatographic dispersion and thus higher
resolution.8 Also the flowrate may have a role in the different selectivity, as a flowrate of 0.2
mL/min was used, in contrast to the 1.0 mL/min flowrate that was used for the measurement
with the PLGel columns and which can potentially reduce the plate height according to Eq. (3),
depending on the effective constants.
-100
400
900
1400
1900
2400
2900
3400
3900
5,4 5,6 5,8 6 6,2 6,4 6,6 6,8 7
Inte
nsi
ty (
nR
IU)
Time (min)
Phenomenex core-shell B109
PLGA 10.5 kDa
PLGA 21 kDa
PLGA Mixture
Chapter 4: Results & Discussion
34
4.2.2 SEC calibration with polymethylmethacrylate
Figure 20: Calibration curve of core-shell B109 column, created with PMMA standards.
Figure 20 shows the calibration curve of the Phenomenex core-shell B109 column with PMMA
standards using the HPLC setup with the RID. As can be seen from the calibration curve, there
is a linear region between the permeation limit and the molar weight of 104.4 g/mol, and a non-
linear region afterwards, which might indicate the exclusion limit in which HDC effects occur.
This HDC region is clearly visible on the calibration curve for this column with the used
standards due to the small particles (98 Å poresize) in the stationary phase. Using a third order
fitting, the PLGA sample sizes were subsequently calculated, and given in Table 3 below.
Table 3: PLGA molar weight calculated according to the PMMA calibration curve fitting
given in Figure 20.
Calibration PLGA 21 kDa PLGA 10.5 kDa
PMMA 600.67 g/mol 5996.6 g/mol
y = -34,306x3 + 131,7x2 - 174,44x + 82,699R² = 0,997
0
1
2
3
4
5
6
7
0,8 0,9 1 1,1 1,2 1,3 1,4
Log
MW
VR (mL)
PMMA calibration curve of Phenomenex core-shell B109 column
PMMA Calibration
Poly. (PMMA Calibration)
Chapter 4: Results & Discussion
35
As can be seen from this table, the calculated molar weight distributions are not in line with
the expected values, as both the PLGA of size of 21 kDa and 10.5 kDa have significant lower
calculated molar weights, even though the samples have molecular weights that fall within the
linear region of the calibration curve. The results do show that the 10.5 kDa PLGA has a higher
molecular mass compared to the 21 kDa PLGA, which is in agreement to the results given in
Table 2 and might be an indication that the 21 kDa PLGA had been substantially decayed. The
unexpected calculated values can also be the consequence of the fact that the PMMA molecules
have a different way in which they fold compared to PLGA in this column with sub-3 µm
particles. A proposed solution would be to use a different set of calibration standards, as this
result shows that PMMA is not capable of providing a calibration curve that is accurate.
4.2.3 SEC Calibration with 3 coupled columns
Figure 21: Calibration curve of core-shell B111, B104 and B109 mixed columns, created
with PS standards.
Figure 21 depicts the calibration curve of the 3 core-shell B111, B104 and B109 mixed
columns. As can be seen from this figure, multiple areas in the curve can be distinguished.
y = -0,9051x3 + 9,6288x2 - 36,563x + 52,733R² = 0,9896
0,00
1,00
2,00
3,00
4,00
5,00
6,00
7,00
2,4 2,6 2,8 3 3,2 3,4 3,6 3,8 4
Log
MW
VR (mL)
PS calibration curve of Phenomenex core-shell B111, B104 and B109 columns
PS Calibration with Slalom
PS Calibration
Poly. (PS Calibration)
Chapter 4: Results & Discussion
36
There is a linear range between a molecular weight of 103.3 g/mol and 104.3 g/mol, after which
an elevation of the calibration curve occurs that continues in a linear fashion between 105.5
g/mol and 106.1 g/mol. These different ranges are caused by the combining of columns, these
stationary phases contain different particle and pore sizes, which has the same effect as a long
mixed column. It is expected that the linear molecular weight ranges are separated using a SEC
mechanism. Above the second molecular weight range, SC is clearly observed which is related
to the stretching of the higher molecular weight polymers in the stationary phase.9 Furthermore,
the data was fitted using a 3th order curve and is fairly precise when leaving out the SC region.
It can be concluded that this set of columns are efficient at separating low-molecular weight
polymers up to a molecular weight of 106.1 g/mol, using a SEC mechanism, but might show
inaccuracy at the elevation area.
4.2.4 Analysis of large polydispersity PS
Figure 22: Chromatogram of large polydispersity PS, blank THF sample and toluene marker,
separated on three Phenomenex core-shell columns (B111, B104 and B109).
Figure 22 depicts the chromatogram of a large polydispersity PS sample, blank and toluene
marker.
-2
0
2
4
6
8
10
12
0 0,5 1 1,5 2 2,5 3 3,5 4 4,5
Inte
nsi
ty (
mA
U)
Time (min)
Phenomenex core-shell B111, B104, B109
Blank
Toluene
PS Broad
Chapter 4: Results & Discussion
37
As can be seen from this chromatogram, the PS sample shows a large distribution which is
characteristic for large polydispersity polymers, and is absent from the blank sample which
indicates that this is indeed the polystyrene sample. It was expected beforehand that due to the
molecular weight distribution overlap with the elevation of the calibration curve at a molecular
weight of 105.2 g/mol, the peak would elute in a non-Gaussian shape. This result shows that
this is not the case, which was unexpected.
To induce this non-Gaussian peak shape, a larger polydispersity polystyrene might need
to be used. Furthermore, it can be seen that during the polystyrene injection, residue of the
toluene sample is injected, which can be solved by running a longer needle wash.
4.2.5 Concluding remarks regarding SEC analysis
From all the SEC measurements performed, it can be concluded that the proposed method with
the respective double columns and core-shell columns are indeed capable of separating the used
PLGA samples, but to be able to make a definite conclusion it is necessary to use undegraded
PLGA building blocks to validate the SEC method and apply these building blocks to prepare
nanoparticles with known compositions. The focus will now go to HDC analysis of polystyrene
samples.
Chapter 4: Results & Discussion
38
4.3 HDC Measurements
Figure 23: Chromatogram of 3-NBS acid separated on an Agilent PL-PSDA Cartridge Type 1
column.
To test the performance of the used HDC stationary phase, column tests have been performed
with 3-NBS acid, uracil and potassium dichromate. Figure 23 depicts the chromatogram of 3-
nitrobenzenesulfonic acid (3-NBS acid) separated on an Agilent PL-PSDA Cartridge Type 1
HDC column, performed under the same conditions as the column test given by the
manufacturer, shown in Appendix 1. As can be seen from this chromatogram, there is a broad
peak visible with a retention time at 3.732 min.
Furthermore, substantial tailing could be observed of several minutes, which is unexpected
when compared to the column performance test supplied by the manufacturer which shows a
sharp peak with no tailing. According to the predicted value of V0 calculated with Eq. (4), from
which the t0 marker retention time is deduced, it can be seen that the expected elution time is
7.2 min. (using a porosity of 0.42). This calculation is an approximation, as the practical elution
time given by the manufacturer is 8.7 min. Nevertheless, it can be concluded that the column
performance test results conducted in this study are not in line with the column test results from
the manufacturer.
Figure 24 and Figure 25 show the injection with uracil and potassium dichromate on the same
column, which are general t0 markers in HDC.
-1
0
1
2
3
4
5
6
7
8
9
0 2 4 6 8 10 12
Inte
nsi
ty (
mA
U)
Time (min)
Agilent PL-PSDA Cartridge Type 1
3-NBS Acid
Chapter 4: Results & Discussion
39
Figure 24: Chromatogram of uracil separated on an Agilent PL-PSDA Cartridge Type 1
column, with a flowrate of 1.6 mL/min and retention time at 3.77 min.
Figure 25: Chromatogram of potassium dichromate separated on an Agilent PL-PSDA
Cartridge Type 1 column, with a flowrate of 1.7 mL/min and retention time at 2.86 min.
As can be seen from both chromatograms of uracil and potassium dichromate injections on the
HDC column, both signals show broad signals with substantial tailing. Uracil has a retention
time of 3.77 min. at a flowrate of 1.6 mL/min, while potassium dichromate has a retention time
of 2.86 min. at a flowrate of 1.7 mL/min, which are early elution times for t0 markers on this
column, calculated under the respective conditions according to Eq. (4).
-50
0
50
100
150
200
250
300
350
400
450
0 2 4 6 8 10 12 14 16
Inte
nsi
ty (
mA
U)
Time (min)
Agilent PL-PSDA Cartridge Type 1
Uracil
-10
0
10
20
30
40
50
60
70
80
0 1 2 3 4 5 6 7 8
Inte
nsi
ty (
mA
U)
Time (min)
Agilent PL-PSDA Cartridge Type 1
Potassium dichromate
Chapter 4: Results & Discussion
40
Figure 23-25 therefore suggest that the column stationary phase had been degraded. A possible
explanation could be that the stationary phase was damaged during column conditioning.
It can be concluded that the available Agilent PL-PSDA Cartridge Type 1 column
could not be used for the analysis of polystyrene analysis. This issue could be resolved by
using a replacement of this HDC column.
Chapter 5: Conclusion
41
5. Conclusion
With the analysis of PLGA on the different SEC columns investigated, insight was gained on
the various column efficiencies. It can be concluded that some columns are indeed more
effective for the separation of the PLGA mixture with labelled 10.5 kDa and 21 kDa molar
weight distributions. Especially the PLGel Mixed double column and the Phenomenex core-
shell column showed good separation results, where both signals in the mixture could be clearly
distinguished. When the retention times of the PLGA samples were related to the calibration
curves prepared with PS or PMMA standards, the PLGel Mixed double column showed
accurate molecular weight results for the 10.5 kDa PLGA sample, while the molecular weight
calculation performed with the Phenomenex core-shell column gave results that were
inaccurate for both 21 kDa and 10.5 kDa PLGA samples. The latter column can therefore not
be used for molecular weight determinations under the conditions that are applied in this study.
The high efficiency of the Phenomenex core-shell column can furthermore be related to the
smaller particles in the stationary phase which decreases the plate height and therefore
increases the resolution.
Moreover, it can be concluded that the PLGA with a molecular weight of 21 kDa had
been degraded, which was confirmed by SEC molecular weight determination measurements
with the PLGel Mixed double column and the Phenomenex core-shell column, while it shows
that the 10.5 kDa PLGA sample still had the original molecular weight distribution. In this
study it is also shown that there is no significant difference between using PS or PMMA as
calibration standards, as the calibration curves resulting from injections with either of the
polymers were similar.
To validate the proposed second dimension SEC methods for the comprehensive
analytical MAnIAC system, HDC was performed with the Agilent PL-PSDA Cartridge Type
1 column, but as shown during the column test with 3-NBS acid, uracil and potassium
dichromate, the performance of this column was insufficient due to the extensive tailing of
peaks and retention times that did not correspond to the specifications of the manufacturer.
Further HDC measurements could therefore not be conducted with this column to validate the
SEC methods which were developed in this study.
Chapter 6: Future Prospects
42
6. Future Prospects
A priority in this study was testing out the right columns and conditions in which polymer
mixtures can effectively be separated by the SEC mechanism. These findings can subsequently
be used in the optimization of the envisaged comprehensive analytical system for the analysis
of polymers. It is expected that especially the application of Phenomenex core-shell columns
are a good choice for this system, due to the high flowrates that can be utilized which facilitates
faster separations in the second dimension. Furthermore, the Phenomenex core-shell column
proved to give the highest resolution of all the tested columns, despite the smaller void volume
of core-shell columns compared to columns with entirely porous particles. For future work, it
would be viable to use the 3 coupled core-shell columns to inject the PLGA samples, and
determine its molecular weight distributions with undegraded PLGA building blocks.
An important stage in the development of the second dimension is the validation of the
method. It was envisaged to perform this validation with HDC, but due to issues with the
stationary phase of the HDC column, this was not performed. In future studies, it is advised to
continue the planned validation with HDC, to see if the method is indeed capable of separating
polymer nanoparticles with known polymer lengths and compositions.
An extension of this work would be investigating the use of RPLC or IEC columns
instead of the SEC separation methods studied in this research. Using RPLC or IEC columns,
other characteristics of polymers like hydrophobicity and charge can be determined.
The next step in the development of the comprehensive analytical system would be
combining both first and second dimension separation methods. To do this, HDC and SEC
conditions need to be optimized for both methods to achieve the best possible performance.
References
43
References
[1] American Chemistry Council. The Basics: Polymer Definition and Properties
https://plastics.americanchemistry.com/plastics/The-Basics/ (accessed May 20, 2016)
[2] Understanding Nano. Nanoparticle Applications and Uses.
http://www.understandingnano.com/nanoparticles.html (accessed May 20, 2016).
[3] DSM Corporate Public Affairs. DSM position on nanotechnologies.
https://www.dsm.com/content/dam/dsm/cworld/en_US/documents/nanotechnology.pdf
(accessed May 20, 2016).
[4] DSM. DSM in Paint & Coatings. https://www.dsm.com/markets/paint/en_US/home.html
(accessed May 21, 2016).
[5] Edam, R. Comprehensive Characterization of Branched Polymers. Ph.D. Dissertation,
University of Amsterdam, Amsterdam, 2013.
[6] Harris, D. C. Quantitative Chemical Analysis, 8th ed.; W. H. Freeman and Company: New
York, 2010.
[7] The Linde Group. High-Performance Liquid Chromatography (HPLC).
http://www.lindeus.com/en/industries/laboratories/analysis_and_measurement/high-
performance_liquid_chromatography/index.html (accessed Jun 9, 2016).
[8] Uliyanchenko, E. Size-exclusion chromatography – from high-performance to ultra-
performance. Anal. Bioanal. Chem. 2014, 406, 6087–6094.
[9] Striegel, A. M.; Yau, W. W.; Kirkland, J. J.; Bly, D. D. Modern Size-Exclusion Liquid
Chromatography, 2nd ed.; Wiley: Hoboken, 2009.
[10] Striegel, A. M.; Brewer, A. K. Hydrodynamic Chromatography. Annu. Rev. Anal. Chem.
2012, 5, 15–34.
[11] Shimadzu. Size Exlusion Chromatography. http://www.shimadzu.com/an/hplc/support/
lib/lctalk/55/55intro.html (accessed May 29, 2016).
[12] Shodex. Calibration Curve for LF-804 (THF Eluent: PS) https://www.shodex.com/en/
dc/06/05/83.html (accessed May 29, 2016).
[13] Eclipse Business Media ltd. Separation Science, HPLC Solutions #101: Estimating
Column Volume. http://www.sepscience.com/Techniques/LC/Articles/842-/HPLC-Solutions-
101-Estimating-Column-Volume (accessed June 24, 2016).
[14] Popovici, S-T.; Kok, W. Th.; Schoenmakers, P. J. Band broadening in size-exclusion
chromatography of polydisperse samples. J. Chromatogr. A. 2004, 1060, 237–252.
[15] Preti, R. Core-Shell Columns in High-Performance Liquid Chromatography: Food
Analysis Applications. Int. J. Anal. Chem. 2016, 2016, 9 p.
References
44
[16] American Chemical Society. 55th Annual Report on Research 2010.
https://acswebcontent.acs.org/prfar/2010/AC7.html (accessed Jun 9, 2016).
[17] Uliyanchenko, E. V., Ultra-performance polymer separations. Ph.D. Dissertation,
University of Amsterdam, Amsterdam, 2012.
[18] Eppert, J. E. HPLC Troubleshooting, 1st ed.; Separation Service Berlin: Berlin, 2003.
[19] Agilent Technologies. PLGel Mixed column specifications. https://www.agilent.com/
en-us/products/gpc-sec/gpc-sec-columns/organic-gpc/plgel-mixed (accessed Jun 28, 2016).
[20] Makadia, H. K.; Siegel, S. J. Poly Lactic-co-Glycolic Acid (PLGA) as Biodegradable
Controlled Drug Delivery Carrier. Polymers 2011, 3, 1377–1397.
[21] Sigma Aldrich. Reference: Polymer Properties https://www.sigmaaldrich.com/content/
dam/sigma-aldrich/docs/Aldrich/General_Information/viscosity.pdf (accessed Jun 5, 2016).
Appendix
45
Appendix 1: Column performance report of Agilent PL-PSDA Cartridge Type 1 column