    Agilent 1100 Series

    Liquid Chromatograph 

    Hardware and Software Installation Checklist

    Hardware / SoftwareUnpacking and inspection of

    shipped materials 

      solvent inlet/outlet

     

    Priming each channel withIsopropanol from solvent bottleto pump inlet

    InstallationNote: Shipping containers shouldnot be opened until an AgilentTechnologies representative is present

    Purpose: To ensure that instrumentsand systems are correctly installedand functioning as designed, in thecustomer's facility. Correctinstallation is the first step inensuring that instruments andsystems operate reliably over anextended lifetime.

    G1310A/11A/12A – Pump

      Section NOT Applicable   Unpack all boxes and place equipment on bench. To prevent injury, get liftingassistance if needed.

    1. SN/FW#_____________________ 

    2. SN/FW#_____________________   Retain shipping containers andmaterial until installation iscomplete and performance is

    verified

    3. SN/FW#_____________________ 

    Customer Responsibilities The customer should ensure that theinstallation site is prepared inaccordance with the specificationscontained in the relevant site preparation document and that thenecessary operating supplies,consumables and usage dependentitems such as vials, paper, solvents,etc. are available. A customerrepresentative should be present atall times during the installation.

      Set up the pump(s)   Notify order fulfillment of any

    discrepancies    solvent inlet/outlet  waste drain

      Unpack and verify conditionand completeness of shipment   remote control (if used)

      relay contacts (if installed)  Check serial numbers and fill in

    installation documentation S/Nfields if necessary

      seal wash tubing(if installed)

      CAN  Verify that power cords are

    correct (must mate with powersockets) and are plugged into power receptacle

      Agilent GPIB/LAN (see note*)   Priming the pump

       prepare solvents

     

    turn on 

    When applicable, installcomputer system and connect all power interconnect cables

    Note: Typical installation time is

    1.5 hours for a system and 30

    minutes for a module.

       purge   purge mobile phase  monitor the pressure

      If a G1315C-DAD or G1365C-MWD is added to an existing1100 system, all modules must be upgraded to firmwarerevision A.06.xx using theLAN/RS-232 FW Update Tool2.00 according to the updatedocumentation.

    If installation time was not sold forany section, check that section asnot applicable.

    G1382A G1376A Capillary Pump

    G1379A Micro Degasser

    G2225A G2226A Nano pump

    Issued: April, 2005 – Rev. 5.0 N/A WAD LPA Product Support Page 1 of 6

    Installation Checklist G1379A Micro Degasser 

      Section NOT Applicable   Section NOT Applicable 

    G1322A/79A– DegasserNote:

    If one section contains multiplechoices for installed Hardware please encircle all modules that are part of the system and fill in thecorresponding Serial Number. 

    1. SN/FW#_____________________   Section NOT Applicable 

    2. SN/FW#_____________________ 1. SN/FW#_____________________ 

      Set up the pump2. SN/FW#_____________________ 

      solvent inlet/outlet  waste drain

      Set up the degasser(s)  remote control (if used)

  • 8/17/2019 Inst Checklist 0405

    2/6

     

    Agilent 1100 Series

    Liquid Chromatograph 

    Hardware and Software Installation Checklist

    5065-9922 Prep. Manual Injector

    Issued: April, 2005 – Rev. 5.0 N/A WAD LPA Product Support Page 2 of 6

      relay contacts (if installed)

     

    CAN  Agilent GPIB/LAN (see note*)   set up the degasser  solvent inlet/outlet   priming each channel with

    Isopropanol from solvent bottle to pump inlet

      Priming the pump   prepare solvents  turn on   purge   purge mobile phase

     

    monitor the pressure 

    G1361A – Prep. Pump

      Section NOT Applicable 

    1. SN/FW#_____________________ 

    2. SN/FW#_____________________ 

      Set up the pump(s)  solvent inlet/outlet  waste drain

     

    remote control (if used)  relay contacts (if installed)  seal wash tubing  CAN  Agilent LAN (see note*) 

      Priming a prep pump  Prime each pump individually  Use the syringe + adapter +

    tubing and connect to the wasteoutlet of the pump`s EMPV

      Turn ON the pump and

    electronically switch the EMPVto waste  Manually suck the solvent into

    and trough the pump  Pump at least 0.5 liter of IPA

    through the pump

    G1328A/B – Manual Injector or

    G1327A - Thermostat. ALS

     

    Section NOT Applicable  G1389A - Thermostat. Micro ALSG1364A/B/C/D + G1330A/B

    SN/FW#________________________ Therm. Fraction Collector   G1368A - Thermostat. WPS  Setup the injector G1378A - Thermostat. Micro WPS

      start/stop cable to pump G2261A - Thermostat. Preparative   base plate & mounting pole   position valve  inlet/outlet connected  sample loop  vent tubes

    G1313A/29A – Autosampler

    G1387A – Micro ALS

    G1367A - Well Plate Sampler

    G1377A - Micro WPS

    G2258A – Dual Loop ALS

    G2260A - Preparative ALS 

      Section NOT Applicable 

    1. SN/FW#_____________________ 

    2. SN/FW#_____________________ 

    3. SN/FW#_____________________ 

      Setup the autosampler(s)  solvent inlet/outlet  waste drain (must be in a

    straight line down, no bends, kinks or curves

      large volume tube(if installed)

      BCD output (if installed)  remote control (if used)  relay contacts (if installed)

     

    CAN  Agilent GPIB/LAN (see note*) 

    The following section applies if aG1330A/B - ALS Thermostat

    is installed together with any of thewell-plate-, autosamplers or fractioncollectors to generate a:

    ALS 

    G1330A/B - ALS Thermostat 

      Section NOT Applicable 

    1. SN/FW#_____________________ 

    2. SN/FW#_____________________ 

    3. SN/FW#_____________________ 

      Setup the ALS Thermostat(s)underneath the module(s) thathas (have) to be thermostatted

    WARNING: Do not connect theALS thermostat cable while the line power is connected to the anyof the modules. This would damage

    the sampler or fraction collector andthe thermostat electronics.  air channel adapter and plate  ALS thermostat cable  ALS thermostat waste drain

    G1316A - Column Compartment

      Section NOT Applicable 

    SN/FW#________________________ 

      Setup the TCC  solvent inlet/outlet  column  waste drain  remote control (if used)  CAN  GPIB (see note*) 

  • 8/17/2019 Inst Checklist 0405

    3/6

     

    Agilent 1100 Series

    Liquid Chromatograph 

    Hardware and Software Installation Checklist

    G1157A/58A/59A/60A/62A/63A   remote control (if used)

    Valves

    Issued: April, 2005 – Rev. 5.0 N/A WAD LPA Product Support Page 3 of 6

      Section NOT Applicable 

    1. SN/FW#_____________________ 

    2. SN/FW#_____________________ 

    3. SN/FW#_____________________ 

      Setup the Valves  Setup the 1100 Series Valve  solvent inlet/outlet

     

    CAN  CAN-DC-In from other 1100

    Module (e.g. G1367A WPS,G1364A/B/C/D FractionCollector or G1361A Prep.Pump) 

    G1383A Column Organizer

      Section NOT Applicable 

    SN/FW#________________________ 

      Setup the column organizer   base plate and organizer plate  rods  traversesadditional holders (if used)

    G1314A – VWD

      Section NOT Applicable 

    SN/FW#________________________ 

      Set up the detector  flow cell

     

    solvent inlet/outlet  waste drain  remote control (if used)  analog output (if used)  relay contacts (if installed)  CAN  Agilent GPIB/LAN (see note*) 

      Verify the operation of the

    detector

     

    analog output (if used)  CAN  Agilent GPIB/LAN (see note*) Lamp

    IntensityPass/Fail

      Verify the operation of thedetectorZero-order

    DeviationWLcalibration Lamp

    IntensityPass/Fail656nm

    DeviationExcitationDeviationG1315A/B/C-DAD / G1365A/B/C-

    MWD

      Section NOT Applicable 

    1. SN/FW#_____________________ 

    2. SN/FW#_____________________ 

      Set up the detector(s)  deuterium lamp

    (if not installed)

      solvent inlet/outlet  waste drain  remote control (if used)  analog output (if used)

     

    relay contacts (if installed)  CAN  Agilent GPIB/LAN (see note*)   Verify the operation of the

    detector

    LampIntensity

    1.2.

    Pass/Fail

    1.2.WL

    calibration 1.2.

    AlphaDeviationBetaDeviation

    G1321A – FLD  Section NOT Applicable 

    SN/FW#________________________

      Set up the detector  solvent inlet/outlet  waste drain

    WLverification Emission

    Deviation

    G1362A – RID

      Section NOT Applicable 

    SN/FW#________________________

      Set up the detectorWARNING: Make sure that all plugs are removed from the portsand the appropriate capillaries(inlet/outlet/recycling) areconnected (maximum backpressure

    of 5 bar for flow cell).  solvent inlet/outlet  solvent recycle  waste drain  remote control (if used)  analog output (if used)  CAN  Agilent GPIB/LAN (see note*) 

      Verify the operation of thedetector

    Diode

    Balance Value

    G1364A/B/C/D – Fraction

    Collector

      Section NOT Applicable 

    1. SN/FW#_____________________ 

  • 8/17/2019 Inst Checklist 0405

    4/6

     

    Agilent 1100 Series

    Liquid Chromatograph 

    Hardware and Software Installation Checklist

    2. SN/FW#_____________________ 

    3. SN/FW#_____________________

    4. SN/FW#_____________________ 

      Setup the fraction collector(s)  solvent inlet/outlet  waste drain  remote control (if used)  relay contacts (if installed)  CAN  Agilent LAN  (see note*) 

    G1390A – Universal Interface Box

    (UIB)

      Section NOT Applicable 

    SN/FW#________________________ 

      Setup the universal interface box

      CAN  Analog In (if used)  GPIO / Analog In (if used)  Relay Contacts (if used)

    Note: Agilent GPIB/LAN connectionis made between one (1) of the Agilent

    1100 Series System modules and the

    Agilent ChemStation PC. If a

    detector is installed connect the

    GPIB/LAN to this module.

    Issued: April, 2005 – Rev. 5.0 N/A WAD LPA Product Support Page 4 of 6

  • 8/17/2019 Inst Checklist 0405

    5/6

     

    Agilent 1100 Series

    Liquid Chromatograph 

    Hardware and Software Installation Checklist

    Software Installation (Un- 1100 LC CheckoutG1323A/B – Control Module

    Issued: April, 2005 – Rev. 5.0 N/A WAD LPA Product Support Page 5 of 6

     

    Section NOT Applicable 

    SN/FW#________________________

      Set up the control module  Connect via CAN to

    available instrument module

    Software Installation (Bundled)

      Section NOT Applicable 

    Software

    Product:_________________________

    SoftwareRevision:________________________

      Start Windows  Start ChemStations -> Add

    Licenses and add the licensenumbers. You find thesenumbers in the 'SoftwareCertificate and RegistrationPacket' envelopes.

      Configure TCP/IP if the

    instrument will be connectedusing LAN.

      Install and Configure IOLibraries if instrument will beconnected using GPIB.

      If necessary install andconfigure AGILENT BootpService if the instrument will beconnected using LAN.

      Use the Configuration Editorand configure your instruments.

      Install and Configure Software

    Add-Ons (if applicable)  If present, install the PC Image

    software and create a backupusing the Backup Solution UserGuide as reference.

    Bundled)  

    Start a run  Section NOT Applicable   Prepare:

    1.   pump(s)SoftwareProduct:_________________________

    2.  online degasser (s)(if present)

    3.  injection device(s)SoftwareRevision:_____________________  

    4.  column compartment5.  detector(s)6.  fraction collector(s)

    (if present)  Start Windows  Install ChemStation using the

    ChemStation Installation guide

    for reference.

    7.  Valve(s) (if present)8.  recording device

     

    Attach results to installationdocumentation

      Check if a printer driver isinstalled on the system

      Fill out instrument logbook  Verify that the regional settings

    are set to English-US  Install and Configure TCP/IP if

    the instrument will be connectedusing LAN.

      Install and Configure IOLibraries if instrument will beconnected using GPIB.

      If necessary install and

    configure AGILENT BootpService if the instrument will beconnected using LAN

      Use the Configuration Editorand configure your instruments.

  • 8/17/2019 Inst Checklist 0405

    6/6

     

    Agilent 1100 Series

    Liquid Chromatograph 

    Hardware and Software Installation Checklist

    Service Review

    Issued: April, 2005 – Rev. 5.0 N/A WAD LPA Product Support Page 6 of 6

    Note: Password is the Registration

    number for the software  Affix any reports generated tothis Checklist

      Sales Order Number  Record in the instrumentlogbook date and time installwas completed

     _________________________

      Customer Service Order (CSO) Number

      Explain Agilent warranty forHardware, Software and return policy for the InstrumentService Center

     __________________________

      Date completed  Explain how to log aninstrument service call  __________________________

      Explain support services

     

    Customer Signature 

    Explain how to use manuals  Advise customer of additional

    training options  __________________________

      Advise the customer of thefollowing useful AgilentWebSites:

      Support Provider Signature

     __________________________  e.g. 1100 Series Firmware(Downloads/patches)

    Exclusive offers:

    www.agilent.com/chem/exclusiveoffers 

    Library -http://www.agilent.com/chem/library

    Education -www.agilent.com/chem/education

    eSeminars -www.agilent.com/chem/eseminars

    Software Status Bulletins and Patches -www.agilent.com/chem/techsupport

    Supplies - www.agilent.com/chem/supplies

    Tech Support -

    www.agilent.com/chem/techsupp

     

    http://www.agilent.com/chem/exclusiveoffershttp://www.agilent.com/chem/libraryhttp://www.agilent.com/chem/educationhttp://www.agilent.com/chem/eseminarshttp://www.agilent.com/chem/techsupporthttp://www.agilent.com/chem/supplieshttp://www.agilent.com/chem/techsupporthttp://www.agilent.com/chem/techsupporthttp://www.agilent.com/chem/supplieshttp://www.agilent.com/chem/techsupporthttp://www.agilent.com/chem/eseminarshttp://www.agilent.com/chem/educationhttp://www.agilent.com/chem/libraryhttp://www.agilent.com/chem/exclusiveoffers