D Determinator Full
-
Upload
thouche007 -
Category
Documents
-
view
227 -
download
1
Transcript of D Determinator Full
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 1
User’s Manual
D-Determinator
Software Version 1.21August 10, 2001
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 2
TABLE OF CONTENTS
INTRODUCTION..................................................................................................................................................4
GENERAL DESCRIPTION ............................................................................................................................................4D-DETERMINATOR CONTROL FEATURES ..................................................................................................................6
HARDWARE SETUP............................................................................................................................................7
NOTES ON SAFETY ....................................................................................................................................................7MOUNTING AND LOCATION .......................................................................................................................................7WIRING NOTES ..........................................................................................................................................................7POWER WIRING.........................................................................................................................................................7ANALOG INPUTS........................................................................................................................................................7DIGITAL INPUTS ........................................................................................................................................................7DIGITAL OUTPUTS .....................................................................................................................................................8FIELD I/O LAYOUT....................................................................................................................................................9
SYSTEM CONFIGURATION............................................................................................................................13
GENERAL ................................................................................................................................................................13SYSTEM FLOW CHARTS AND CONTROL PANEL KEYS ..............................................................................................14ENTERING CONFIGURATION MODE .........................................................................................................................15SELECT LANGUAGE .................................................................................................................................................15ANALOG SENSOR INPUT CONFIGURATION ...............................................................................................................16
Input Configuration Flow Chart ........................................................................................................................16Select sensor type of input 1...............................................................................................................................16Configuration of conductivity sensor (Analog Input 1)......................................................................................17Configuration of pH-sensor (Analog Input 1)....................................................................................................19Configuration of RedOx-sensor (Analog Input 1) ..............................................................................................21Configuration of ClO2-sensor (Analog Input 1).................................................................................................21Configuration of Direct (Analog Input 1) ..........................................................................................................22Select sensor type of input 2...............................................................................................................................23Configuration of temperature sensor (Analog Input 2)......................................................................................23Configuration of RedOx-sensor (Analog Input 2) ..............................................................................................24
EQUIPMENT CONFIGURATION..................................................................................................................................25Setting Equipment Mode (On/ Off).....................................................................................................................26Select flowmeter .................................................................................................................................................26Select pump type.................................................................................................................................................27Flow Rate Input / Teach In.................................................................................................................................27Control of product flow......................................................................................................................................29
OPERATION MODE ..................................................................................................................................................30Select Operation Mode.......................................................................................................................................30Analog Regulator ...............................................................................................................................................30Volume fix ..........................................................................................................................................................32Volume fix + secondary Control ........................................................................................................................32Dependant ..........................................................................................................................................................33Setpoints for Initial Charge and Start-Up..........................................................................................................34Return to Configuration Menu ...........................................................................................................................34
READING / RESETTING TOTAL CONSUMPTION.........................................................................................................34NETWORK SETTING .................................................................................................................................................35
Changing the Password .....................................................................................................................................35Input / Output Test .............................................................................................................................................36Resume Operation..............................................................................................................................................38
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 3
OPERATING INSTRUCTIONS.........................................................................................................................39
OPERATION OF D-DETERMINATOR ..........................................................................................................................39AUTOMATIC DOSING................................................................................................................................................41
Analog Regulator ...............................................................................................................................................41Volume fix + secondary Control ........................................................................................................................41Volume fix ..........................................................................................................................................................42Dependant ..........................................................................................................................................................42
DOSING AN INITIAL CHARGE/ START-UP.................................................................................................................42MANUAL DOSING ....................................................................................................................................................43OPERATION AS A "SLAVE".......................................................................................................................................45NETWORK OPERATION.............................................................................................................................................45
TROUBLESHOOTING.......................................................................................................................................46
ALARMS AND WARNINGS........................................................................................................................................46Alarm messsages ................................................................................................................................................46Warning messages..............................................................................................................................................48
SYSTEM NOT FUNCTIONAL .....................................................................................................................................49
EXAMPLE............................................................................................................................................................50
APPENDIX A: SPECIFICATIONS ................................................................................................................................52APPENDIX B: ELECTRICAL CONNECTIONS...............................................................................................................53APPENDIX C: MENU SCHEMES ................................................................................................................................54
Operation Menu .................................................................................................................................................54Configuration Menu ...........................................................................................................................................55Input Configuration, Submenu to Configuration ...............................................................................................56Equipment Configuration, Submenu to Configuration.......................................................................................57
APPENDIX D: PARAMETER LIST ..............................................................................................................................58APPENDIX E: MOUNTING DIAGRAM ........................................................................................................................62APPENDIX F: SPECIFIC CONDUCTIVITY DIAGRAMS ..................................................................................................63
Conductivity for different solutions....................................................................................................................63Relationship between RedOx potential and ClO2-Concentration ......................................................................64
GLOSSARY..........................................................................................................................................................65
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 4
Introduction
General DescriptionThe D-Determinator is part of a modular system solution for the dosing of cleaning products into cleaning application (i.e.CIP, bottlewashers or other cleaning process). It offers a high level of flexibility: The D-Determinator can operate by itself ornetworked for a multi-zone application.
• The concentration of cleaning solution can be controlled via a fixed volume of product per time or by measuringthe conductivity, pH or the RedOx potential and controlling to a desired setpoint.
• Up to two additional products can be added either as fixed volume per time or proportional to the volume of themain product.
• A separate and typically larger initial charge pump can be configured to add the main product faster.
• Up to four D-Determinators can be networked together, via CAN-Bus, using the same equipment controllingcleaning solution being distributed to multiple dosing points.
The process equipment that can be monitored and controlled by a D-Determinator unit is shown below. Not all of theequipment shown below is required for every dosing application. To improve material tracking functionality additional flowtransmitters can be added to monitor consumption. Low level sensors can be added to check the level of product in the supplytanks or drums.
All these process signals can be connected to the D-Determinator, but the only equipment that is required are the dosingpumps.
* * *
Main Product
DosingPump
(Equip. 1)
InitialChargePump -Optional
(Equip. 4)
Day Tank orEquipment to becleaned
DosingPump
(Equip. 2)
Additive 1
DosingPump
(Equip. 3)
Additive 2
- Pulse flow transmitter for totalizing the product flow
LS
FT
FT FT
LS LS
FT
LS - Low Level Switch (Depletion alarm)
TT
CT
TT
CT - Conductivity, pH, Redox or ClO2 transmitter can be connected
- Temperature transmitter can be connected to perform temperaturecompensation on the conductivity signal.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 5
The D-Determinator system is comprised of a wall mountable control unit with a separate access area for the field connections.The control panel has 4 keys and a LCD display with 2 rows by 16 characters per row. All activities of the dosing system arecontrolled and monitored from the LCD display and the attached external devices. The control system is configured using thecommands and input screens on the LCD display.
The D-Determinator has four electronic AC relay outputs for controlling four pieces of dosing equipment (dosing pumps orvalves). For each dosing item, there is one ‘power’ digital output driving the pump or, in "Slave"-Mode the valves. Fouradditional relay outputs control the stroke pulses if stroked dosing pumps are used. TheD-Determinator’s dosing operations can be activated and inhibited from an external device through the trigger release input."Initial Charge" and "Start-Up" dosing can also be started by an external device through separate trigger inputs. Optionally,the D-Determinator system can be networked to a computer running configuration and monitoring software that operatesthrough the CAN-Bus interface.
More than one D-Determinator may be connected together using the same process equipment to control different stages of amachine. In this case one D-Determinator is "Master" and controls all dosing equipment and one part of the machine. Theother D-Determinator (s) act as slaves and control valves from the transport line to "their" part of the machine.
When a system with a single D-Determinator is running in Automatic Mode, the D-Determinator controls the dosingequipment per the configuration enter into the controller. A system composed of more than one D-Determinator running inAutomatic Mode requires a CAN-Bus -Interfaces. The D-Determinator configured as the “Master“ controls the dosingequipment and all "Slave" nodes make request for product to the “Master” and wait for a response.
* * *
Master Slave 1 Slave 2 Slave 3
CAN-Bus
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 6
The possible operation modes for a D-Determinator are as follows:
No Release : No dosing activity possibleAutomatic : Dosing automaticallyInitial Charge : Dosing an initial chargeStart-Up : Dosing a start-up chargeManual : Dosing a manually set charge of product
D-Determinator Control Features
• Controls dosing equipment for one main product and up to two supplementary products in Automatic or ManualMode.
• Monitors daily and total consumption for each cleaning product via a pulse flowmeter or by counting the time thedosing pump is running. The pulse flowmeter option provides a more accurate consumption amount, but it requiresadditional process equipment.
• Triggers alarm messages and halts current processing step when process alarms occur.
• Systems can be networked through an optional CAN-Bus interface.
• The configuration menus are password protection.
• Several different operating modes for controlling the dosing pumps.
• Several different operating modes for controlling the cleaning solution’s concentration.
• An additional high capacity pump can be connected to reduce the time required to complete an initial charge orstartup charge.
This document is describing version 1.21 of the D-Determinator hardware and software.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 7
Hardware Setup
Notes on SafetyTo ensure that the controller functions perfectly and has a long, useful life, the user must observe the instructions in thismanual and must proceed according to the conditions of use. Commissioning and maintenance personnel must have beentrained and authorized to carry out such activities. Appropriate measures must be taken to prevent unintended actuation andcausing detrimental influences on the process. The applicable prevention and safety regulations for electrical devices must beobserved during operation, maintenance and repair of the controller.
A qualified electrician must make all electrical connections. Always disconnect from power supply before opening the cover.Warning: Danger of electrical shock inside.
Electrical components should not be exposed to water or moisture. If this occurs by accident, a qualified electrician mustservice the unit.
A qualified electrician following all the local and country electrical codes should install the D-Determinator.
Mounting and locationThe D-Determinator unit should be mounted on a clean, vertical surface and kept away from heat (<50°C), moisture andvibrations. As shown in the mounting diagram (Appendix E), there are six mounting holes to mount the D-Determinator unitto a wall - three on each side on the back wall of the plastic box. When you have located the place for the control unit, mountthe box securely in place.
See flow sensors manual for proper mounting of the desired flow sensor.
See conductivity sensors manual for proper mounting of the desired conductivity sensor.
If other sensors are used such as pH or RedOx verify they are installed properly by checking their installations manuals.
Wiring notesFor the electrical wiring of the components, pay attention to the maximum cable lengths between the D-Determinator unit andthe other components (refer to the technical manuals of the components). Connect wiring as shown in wiring diagram(Appendix B).
Power WiringMake sure the power is OFF which connecting it to the D-Determinator. Verify the power jumper is in the proper location(115VAC or 230VAC) before connecting power to the D-Determinator. Make sure that all attached equipment is inconformance with the electrical specifications (Appendix A).
Analog InputsAnalog Sensor: There are two analog input channels available on the D-Determinator. They can be configured forconductivity (typically), RedOx, ClO2, temperature, pH or direct.
Flow Sensor: The three pulse flow sensors channels. These flow sensors are optional equipment and can be disabled duringconfiguration if not connected. In this case no direct measurement of product consumption is possible; the product flow willbe calculated from the characteristics of the dosing pump. The product consumption’s will also be calculated from the pumpcharacteristics.
Digital Inputs
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 8
These digital inputs are potential free digital input signals. All input clamps also feature 0 volt and 24 volt access.
Low Level Switches: Connect level switches for the empty level. An ‘OFF’ signal (open contact) refers to an empty tank; a‘ON' signal (closed contact) refers to a tank that is not empty. A broken wire or a faulty connection will also prompt an alarm.If this contact is opened while product is running, an immediate alarm will be produced and the dosing pump will be stopped.If there isn’t and actual low level switch in the product drum then these channels need to have a manual jumper install acrossthe contact to allow the dosing pump to run. If this jumper is install there is a good possibility that the pump will run when thechemical drum is empty. This may cause damage to the dosing pump. It is recommended that a low level switch be installedon the pickup lance of the chemical and wired to these inputs for safety.
Release Trigger. The Release Trigger can be connected to a switch or to an external device, for remote triggering of thesystem. The system will run in automatic mode when the external contact is closed.
Initial Charge/ Start-Up. These inputs can be connected to a switch or to an external device, for remote the starting of aninitial charge (or start-up) dosing. The system will start when the external contact is closed.
Machine Counter. This input is used to dosing product based on an external signal that represents machine activity. Such asa rack counter for a bottlewasher or a keg counter for a keg washer. The cleaning product will be dosed proportional to therate the machine is running per this signal.
Digital Outputs
Connect the 4 dosing pumps to the respective output ports. If you are using a stroke pump, connect the pump also to thecorresponding Stroke output port. The Stroke outputs are the optical relays that are labeled SxC and SxE, where x is thedosing pump number 1 to 4. These channels have a maximum voltage of 24 volts DC. There are also control relays that thedosing pumps can be wired to that can handle higher voltages. These control relays are labeled PxA and PxB where x is thedosing pump number 1 to 4. The PxA relay contacts are for 24 volts DC and the PxB relay contacts are for supply voltage(either 115VAC or 230 VAC). See the wiring sketches below for more details.
The D-Determinator system provides two other relay outputs to notify the operator when the system is in alarm and the systemis running. The alarm output can be used to trigger an external device, such as a lamp or a horn. The running output signifiesnormal operation, and can be attached to an external device, also.
For the maximum output ratings, see the Specifications sheet (Appendix B)
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 9
Field I/O Layout
AI AI PI PI PI DI DI DI DI DI DI DI DO DO DO DO DO DO
CA
N-B
us
CA
N-B
us
Ana
log
Inpu
t 1
Ana
log
Inpu
t 2
Inpu
t Flo
w S
enso
r 1 (P
ulse
)
Inpu
t Flo
w S
enso
r 2 (P
ulse
)
Inpu
t Flo
w S
enso
r 3 (P
ulse
)
Leve
l Ala
rm 1
Leve
l Ala
rm 2
Leve
l Ala
rm 3
Sta
rtup
Trig
ger
Initi
al C
harg
e T
rigge
r
Run
Trig
ger
Rac
k In
put (
Mac
hine
Spe
ed)
Ala
rm R
elay
Run
Rel
ay
Pum
p 1
Str
oke
Rel
ay
Pum
p 2
Str
oke
Rel
ay
Pum
p 3
Str
oke
Rel
ay
Pum
p 4
Str
oke
Rel
ay
CH CH + + + + + Alm1 Run S1C S2C S3C S4CPE PE AI1 AI2 FI1 FI2 FI3 LA1 LA2 LA3 Start Init Trig Rack1 PE PE PE PE PE PECL CL ⊥ ⊥ ⊥ ⊥ ⊥ GND GND GND GND GND GND GND Alm1 Run S1E S2E S3E S4E
PWR PWR PWR PWR DO DO DO DO DO DO DO DO PWR PWR PWR PWR
Ava
ilabl
e 24
VD
C P
ower
Ava
ilabl
e 24
VD
C P
ower
Ava
ilabl
e 24
VD
C P
ower
Ava
ilabl
e 24
VD
C P
ower
Pum
p 1
DC
con
trol
Pum
p 1
AC
con
trol
Pum
p 2
DC
con
trol
Pum
p 2
AC
con
trol
Pum
p 3
DC
con
trol
Pum
p 3
AC
con
trol
Pum
p 4
DC
con
trol
Pum
p 4
AC
con
trol
Sup
ply
Vol
tage
Pow
er (
115V
AC
or
230V
AC
)
Sup
ply
Vol
tage
Pow
er (
115V
AC
or
230V
AC
)
Sup
ply
Vol
tage
Pow
er (
115V
AC
or
230V
AC
)
Sup
ply
Vol
tage
Pow
er (
115V
AC
or
230V
AC
)
Det
erm
inat
or In
put P
ower
+ + + + P1A P1B P2A P2B P3A P3B P4A P4B L1 L1 L1 L1 L1PE PE PE PE PE PE PE PE PE PE PE PE PE PE PE PE PE
GND GND GND GND P1A P1B P2A P2B P3A P3B P4A P4B N N N N N
Below is a typical wiring diagram for an analog input channel when thetransmitter is providing the current loop power.
Below is a typical wiring diagram for an analog input channel when the D-Determinator is providing the current loop power.
+
AI1
⊥
Field transmitter isproviding the currentloop power
+
AI1
⊥
D-Determinator isproviding the currentloop power
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 10
Below is a typical wiring diagram for a pulse flowmeter input channel when the D-Determinator is providing the loop power.
Below is a typical wiring diagram for the digital outputchannels when the D-Determinator is providing the devicewith 24VDC power.
Below is a typical wiring diagram for the digital outputchannels when the D-Determinator is providing the devicewith 115 VAC or 230 VAC power depending on the inputpower to the D-Determinator.
If the field device already has power then the only wire to the relay contacts. Connect to the PxA contact for 24VDC and PxBcontacts for AC voltage up to 230VAC.
Below is a typical wiring diagram for the digital output Below is a typical wiring diagram for the digital output
+
FI1
⊥
Pulse Flowmeter
VDC
OUT
GND
+24V
PE
GND
ControlDevice(Pump
orValve)
PxA
PE
PxA
Jumper
PxB
PE
PxB
ControlDevice(Pump
orValve)
L1
PE
N
Jumper
PxA or PxB
PE
PxA or PxB
ControlDevice(Pump
orValve)
Externalpowersupply
x is 1, 2, 3, or 4depending on thedigital outputchannel number.
x is 1, 2, 3, or 4 depending on the digital output channel number.
x is 1, 2, 3, or 4 depending on the digital output channel number.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 11
channels when the D-Determinator is providing thecontrol device with 24VDC power for the stroke pumpchannels.
channels when an external power supply is providing thecontrol device with 24VDC power for the stroke pumpchannels.
+24V
PE
GND
ControlDevice(Pump
orValve)
SxC
PE
SxE
Jumper
x is 1, 2, 3, or 4depending on thedigital outputchannel number.
SxC
PE
SxE
ControlDevice(Pump
orValve)
External24 V power
supply
x is 1, 2, 3, or 4 depending on the digital output channel number.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 12
The left most connections on the wiring panel are for the CAN-Bus connections. The CAN-Bus is a daisy chain network thatshould be connected like the diagram below.
MasterD-Determinator
CH
PE
CL
CH
PE
CL
Slave 1D-Determinator
CH
PE
CL
CH
PE
CL
Slave 2D-Determinator
CH
PE
CL
CH
PE
CL
Slave 3D-Determinator
CH
PE
CL
CH
PE
CL
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 13
System Configuration
GeneralThe D-Determinator system can automatically control the chemical concentration of the cleaning solution by feeding the necessaryproduct into the machinery. To accomplish this, the D-Determinator monitors 2 analog sensors and controls up to 4 pumps orvalves.
Sensors: Sensor 1: measures the cleaning solution’s conductivity, pH, RedOx, ClO2 ordirectly measures any analog transmitter (current loop)
Sensor 2: measures temperature for compensation of sensor 1, simply measuretemperature or a second RedOx sensor can be connected.
Pumps: Equipment 1: doses main product
Equipment 2: doses additional product 1
Equipment 3: doses additional product 2
Equipment 4: pumps main product during Initial Charge or Startup
Equipment 1 can be configured to perform a time constant dosing ("Volume Fix") or to control the concentration of cleaningsolution to an analog setpoint (“Analog Regulatory”). To pump the higher amount of the main product during an initial chargeoperation a larger pump can be installed and connected to Equipment 4. This pump will only be used during Initial Charge andStart-Up operations. While the system is running in Automatic Mode, dosing will be performed automatically until the TriggerRelease input is OFF (“open contact”) requesting the D-Determinator to stop automatic control.
Equipment 2 and 3 can be configured to deliver product on a time constant dosing ("Volume Fix") or to control to a ratio of themain product dosed, this mode is call Dependant. In dependant mode the amount of product 2 and 3 that will be dosed is based onthe ratio of the consumption amounts, main product to product 2 and/or product 3. The product may or may not be dosed at thesame time depending on the minimum feed amounts for products 2 and 3.
When a process alarm (i.e. low level in product drum) is generated the dosing pumps are stopped. There are several alarms thatwill cause the dosing to stop and there are others that are only warnings, they are listed in Appendix G.
The configuration menus and consumption resets are password protected, so only authorized individuals can perform these tasks.When inappropriate values are entered in the Configuration Menu, or a setting is not possible, the D-Determinator system willinform the user.
While the system is being configured, the D-Determinator system will not perform any dosing - all of the system outputs areblocked and the field signals are set OFF (open contact). When the Configuration Menu is exited, the system is immediately readyto perform the desired dosing operation.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 14
System Flow Charts and Control Panel KeysThe four system flow charts - Operation Menu and Configuration Menu (includes two Submenus) (see Appendix C) - are the roadmaps that will guide anyone through the D-Determinator operations and configurations screens. Whether you want to configurethe system to run automatically, activate a manual dosing, or view product consumption totals, the flow charts will show you thepath you must take to accomplish your task.
The D-Determinator system commands, which appear on the LCD display and rotate in a looping fashion. Depending on whichkey you press on the control panel in different operator screens will be displayed depending on which area of the D-Determinatorthe operator is working.
The two control panel keys, Select and Enter, are indicated on the flow charts by ‘S’ and ‘E’, respectively. These keys generallyrepresent the two different options which are available from any given display screen in the D-Determinator Control system. Withthe Enter key, you can also generally proceed through the system and view system settings, without changing the data.
SELECT SELECT: moves to the next menu item but does not start function.
ENTER ENTER: confirms last entry and continues.
The two arrow keys on the control panel are generally used to make configuration selections and set numeric entries:
ò Down arrow: decreases number in digit field ortoggles between ‘Y’ and ‘N’ in Yes / No field,(field values are pre-set with zero entries).
ð Horizontal arrow: moves cursor to the next digit field,(or pre-sets fields with preceding entries).
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 15
Entering Configuration ModeWhen the control unit is turned on for the first time, the start-up screen will read:
Configuration Enter
As shown on the Operation Menu flow chart (bottom left corner), press the Enter key to confirm, and begin Configuration.
Enter Passwordxxxx
The system requires that you enter a password in order to have access to configuration capabilities. Enter the four-digit password,using the down arrow key and the horizontal arrow key. Press the Enter key when the password is completed and the D-Determinator will display the Configuration Menu – for quick references see the Configuration Menu flow chart in Appendix A.
The general setup of the control system will now be configured, product consumption totals can be read and reset here, thepassword may be changed and the inputs and outputs can be tested.
Referring to the Configuration Menu flow chart (Appendix A), use the keys on the control panel to respond to the information onthe LCD display (see System Flow Charts and Control Panel Keys).
Select languageThe language of the user interface can be selected to suit the personal needs of the user/ operator.
Language Setting Enter
If this menu item is selected by hitting ENTER, a choice of four languages is presented
Sel. Language(E/D/F/Sp): E
The following languages are integrated and can be selected:E - EnglishD - GermanF - FrenchSp - Spanish
Use the down arrow to loop through the language selections.
After selection of the language, the actual type and version of the software is displayed.
D-DeterminatorVersion 1.21
The Default Password is 9999.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 16
Analog Sensor Input ConfigurationThe first screen in the Configuration Mode is:
Sensor Config. Enter
As shown in the Control Configuration flow chart, press the Enter key, to confirm and proceed to Sensor Configuration.
Input Configuration Flow Chart
Referring to the Input Configuration flow chart (Appendix C), use the keys on the control panel to respond to the information onthe LCD display (see System Flow Charts and Control Panel Keys).
Select sensor type of input 1
Select sensor1: Conduct.
The main sensor input (analog input 1) can be connected to different types of sensors. Press the down arrow key, to toggle betweenthe following choices:
Menu Selection Description
Conduct. : Conductivity sensor, different options for displayed unit and range as well as internal or externaltemperature compensation can be configured.
pH : The scale of the pH-sensor can be set freely. The sensor calibration can be made automatically.
RedOx : The scale of the RedOx-sensor can be set freely
ClO2 : The full scale range of the RedOx sensor can be set freely.
direct : The direct display of the input current can be in "mA" or in "%" of full scale.
Sel. Current Min1: (0/4mA): 4
In any case the minimum current of the sensor has to set. A selection of 0 mA or 4 mA is possible depending whether a "live zero"(4-20mA) is used or not (0-20mA)
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 17
Configuration of conductivity sensor (Analog Input 1)Pressing the SELECT key can leave every step in configuration. In this case D-Determinator switches back to the "Select Sensor“screen.
Cond. Sel. Unit1: mS
The conductivity sensor input signal (analog input 1) can be displayed in various ways. The direct display in "mS" or "µS" ispossible as well as an indirect display in terms of concentration of a product. to different types of sensors. Press the down arrowkey, to toggle between the following choices:
Menu Selection Description
mS / µS : Conductivity displayed in physical units, either mS or µS. The maximum scale has to be set also("Cond. Scale max.", see below).
NaOH : The concentration of NaOH is displayed. The measured conductivity is transformed into aconcentration value via a stored calculation diagram. The maximum scale has to be set also.
KOH : The concentration of KOH is displayed. The measured conductivity is transformed into aconcentration value via a stored calculation diagram. The maximum scale has to be set also
H3PO4 : The concentration of H3PO4 (Phosphorous acid) is displayed. The measured conductivity istransformed into a concentration value via a stored calculation diagram. The maximum scale has to beset also
H2SO4 : The concentration of H2SO4 (Sulfuric acid) is displayed. The measured conductivity is transformedinto a concentration value via a stored calculation diagram. The maximum scale has to be set also
HNO3 : The concentration of HNO3 (Salpetric acid) is displayed. The measured conductivity is transformedinto a concentration value via a stored calculation diagram. The maximum scale has to be set also
curve : A free concentration is displayed. The measured conductivity is transformed into a concentration valuevia a stored calculation diagram; this diagram is set by the user ("Free Cond. Curve", see below).
Cond. Scale max.1: 2000 mS
The maximum scale of the sensor (analog input 1) is set in a range from 1 to 2000 of the units selected above. This is thedisplayed value when the process sensor is reading full scale or a 20mA current value.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 18
Below are the two screens that the operator can use to enter a free conductivity curve. Relating conductivity to concentrationpercent.
Free Cond. Curve1-1:0500mS=0,00%
Free Cond. Curve1-1:0500mS=0,75%
If the conductivity has to be displayed in terms of a user specified concentration characteristic, data can be entered. For up to 8points of a characteristic (called 1-1...1-8) the conductivity in µS / mS and the corresponding concentration. The characteristiccurve will start with the data enter in to the 1-1 parameter and end with the data entered in the 1-8. When the conductivity valuehas been entered, hit the ENTER key to enter the corresponding concentration. When the concentration value has been enteredpress the ENTER key to advance to the next point on the curve.The relationship between concentration and conductivity has to be a uniformly rising function. Therefor the default values for thefirst point is "0 µS / mS" and "0%". Setting the following points the minimum values for both concentration and conductivity willbe the corresponding value of the previous point. The default values for these points are the values of the previous point. That is,if an additional data point for a characteristic has to be entered, some points will have to be shifted.
It is possible to see all the data points by pressing ENTER multiple times.
Cond.:Temp.Comp.1: (Y/N): N
If no temperature compensation is needed, this option can be set inactive (Select ‘N’). In this case D-Determinator switches back tothe beginning screen of "Select sensor / 1: ...".
If the conductivity transmitter cannot perform the temperature compensation, this option should be activated (Select ‘Y’). In thiscase the second sensor must be a temperature sensor. So, the configuration of sensor 2 will allow no other option besidestemperature.
Cond. Temp.Coeff1: 9,99 %/°C
For temperature compensation two values must be entered: the temperature coefficient and the reference temperature. Thetemperature coefficient describes the change of measured conductivity in conjunction with a change in temperature. The data sheetof the installed conductivity sensor should have a value for it’s temperature coefficient.
Cond. Ref.Temp.1: 25,0 °C
The reference temperature defines, at which temperature the conductivity has been calibrated. The data sheet of the conductivitysensor should have a value for the reference temperature.
A positive temperature coefficient describes a rising measured conductivity with rising temperature; the measured value will bereduced to give the correct reading. A negative temperature coefficient describes a falling measured conductivity with risingtemperature. The example diagram below shows different curves for both positive and negative temperature coefficient.
Modify the conductivity
Enter Key
Modify the concentration %
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 19
Select Sensor1: pH
Configuration of pH-sensor (Analog Input 1)
Navigate in the configuration menu until you get to
Pressing the SELECT key leaves all the step in the sensor configuration unchanged. In this case D-Determinator switches back tothe "Select Sensor” screen. Pressing the ENTER key proceeds with the sensor configuration.
Sel. Current Min1: (0/4mA): 4
In any case the minimum current of the sensor has to set. A selection of 0 mA or 4 mA is possible depending whether a "live zero"(4-20mA) is used or not (0-20mA)
Effect of temperature compensation
0 mS
20 mS
40 mS
60 mS
80 mS
100 mS
120 mS
140 mS
160 mS
0 °C 10 °C 20 °C 30 °C 40 °C 50 °C
Temperature
-2%/°C
+ 0,5%/°C
corrected
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 20
pH- Scale min.1: 00,0
The low end scale of the sensor (analog input 1) is set in a range from 0,0 to 14,0. This is the pH-value necessary to produce theminimum output current (0 mA or 4 mA) by the pH transmitter.
pH- Scale max.1: 14,0 pH
The maximum scale of the sensor (analog input 1) is set in a range from 0,0 to 14,0. This is the pH value necessary to produce20mA output current by the transmitter.
Calibrate-Sensor1: (Y/N): Y
If no calibration is needed, this option can be refused ("N"). In this case D-Determinator switches back to the "Select Sensor”screen. If the sensor has to be calibrated follow the pH sensor supplier’s instructions and enter the values into the D-Determinator.
Calibrate pH=71: 07,1 [Enter]
The first step is calibration at pH = 7, the sensor has to emerged into a pH = 7-solution. The D-Determinator will show themeasured pH value in momentary calibration. If the read out is stable, press ENTER the current measured value will be referred toas 7,0 pH.
Calibrate pH=4/101: 04,1 [Enter]
The second step is calibration at pH = 4 or at pH = 10, the sensor has to emerged into a pH = 4 or pH = 10-solution. The D-Determinator will show the measured value in momentary calibration. If the read out is stable, press ENTER the current measuredvalue will be referred to as 4,0 pH or 10,0 pH. D-Determinator will switch automatically between these values depending whetherthe second calibration value is higher or lower than a pH = 7 value.
Note: On the D-Determinator the "pH-Scale min." should be below pH = 7 and "pH-Scale max." should be abovepH = 7. If calibration is done with pH = 4, then "pH-Scale min." should be below pH = 4. If calibration is done withpH = 10, then "pH-Scale max." should be above pH = 10. Else calibration will give false values. Make sure that the pH= 4, 7, 10 solutions are correct, else calibration will give false values.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 21
Select Sensor1: RedOx
Select Sensor1: ClO2
Configuration of RedOx-sensor (Analog Input 1)
Navigate in the configuration menu until you get to
Sel. Current Min1: (0/4mA): 4
In any case the minimum current of the sensor has to set. A selection of 0 mA or 4 mA is possible depending whether a "live zero"(4-20mA) is used or not (0-20mA)
Pressing the SELECT-key can leave every step in configuration. In this case D-Determinator switches back to the "Select Sensor”screen.
RedOx-Scale min.1: -1000 mV
The low end scale of the sensor (analog input 1) is set in a range from -1000 mV to 1000 mV. This is the RedOx potentialnecessary to produce the minimum output current (0 mA or 4 mA) by the transmitter.
RedOx-Scale max.1: 1000 mV
The maximum scale of the sensor (analog input 1) is set in a range from -1000 mV to 1000 mV. This is the RedOx potentialnecessary to produce 20mA output current by the sensor.
Configuration of ClO2-sensor (Analog Input 1)
Navigate in the configuration menu until you get to
Sel. Current Min1: (0/4mA): 4
In any case the minimum current of the sensor has to set. A selection of 0 mA or 4 mA is possible depending whether a "live zero"(4-20mA) is used or not (0-20mA)
Pressing the SELECT-key can leave this step in configuration. In this case D-Determinator switches back to the "Select Sensor”screen.
ClO2-Scale max.1: 200,00 ppm
The maximum scale of the sensor (analog input 1) is set in a range from 0 ppm to 200 ppm. This is the ClO2 concentrationnecessary to produce a full scale or 20mA output current by the transmitter.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 22
Select Sensor1: direct
Configuration of Direct (Analog Input 1)
Navigate in the configuration menu until you get to
Sel. Current Min1: (0/4mA): 4
In any case the minimum current of the sensor has to set. A selection of 0 mA or 4 mA is possible depending whether a "live zero"(4-20mA) is used or not (0-20mA)
Pressing the SELECT-key can leave this step in configuration. In this case D-Determinator switches back to the "Select Sensor”screen.
Direct Display1: (mA/%) mA
If the input current has to be displayed directly, the option "mA" should be selected.
Press the down arrow key, to toggle between the following choices:
mA : If the input current is to be displayed in mA with at resolution of 0,02 mA. With the minimum valuebeing set in the previous menu selection (“Sel. Current Min.”) and a maximum value of 20mA.
% : If the input current is to be displayed as % of the input range selected with a resolution of 0,1%. Theinput range may be 0-20mA or 4-20mA depending on the "Sel. Current Min." value enteredpreviously. An example would be if the analog input 1 were measuring a current of 8mA, 25,0%would be displayed for a 4-20mA range or 40,0% for a 0-20mA range.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 23
Select sensor type of input 2
Select sensor2: Temp.
The second sensor input (analog input 2) can be connected to different types of sensors. Press the down arrow key, to togglebetween the following choices: This sensor may be fixed depending on what was configured for analog input 1. Such as iftemperature compensation was selected on the conductivity sensor on input 1 then this input must be a temperature sensor.
Temp. : The full-scale range of the temperature sensor can be set freely. This option is chosen automatically ifthe main sensor is set to be a conductivity sensor with additional temperature compensation. In thiscase, the options "RedOx" and "off" cannot be selected.
RedOx : The scale of the RedOx-sensor can be set freely. This option is only possible if the main sensor is setto be a ClO2
sensor or a RedOx Sensor.
off : If no second sensor is used, this input can be deactivated. No additional settings have to be made.
Sel. Current Min2: (0/4mA): 4
If the options "Temp." or "RedOx" are selected, the minimum current of the sensor has to be set. A selection of 0 mA or 4 mA ispossible depending whether a "live zero" (4-20mA) is used or not (0-20mA)
Configuration of temperature sensor (Analog Input 2)Pressing the SELECT-key can leave this step in configuration. In this case D-Determinator switches back to the "Select Sensor”screen.
Temp.-Scale max.2: 150,0
The maximum scale of the sensor (analog input 2) is set in a range from 0°C to 150°C. This is the temperature necessary toproduce 20mA output current by the temperature transmitter.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 24
Configuration of RedOx-sensor (Analog Input 2)Pressing the SELECT-key can leave every step in configuration. In this case D-Determinator switches back to the "Select Sensor”screen.
Secondary Contr.2: (Y/N): Y
If two RedOx sensors are used, the measured value of the second sensor can be used to supervise the first sensor. The first sensorwill always be the leading sensor regarding the dosing operation. If the "secondary control" option is activated ("Y"), the alarmthresholds can be set at which the D-Determinator will indicate an error when the difference in the values from each sensor exceeda configured amount. If the first sensor is a ClO2 sensor, the output signal of the RedOx sensor will be computed to give a ClO2
concentration. The deviation alarms will be set in terms of a ClO2 concentration (ppm).
Deviation min.2: -1000 mV
The minimum deviation between the two sensors (analog input 2 is lower than analog input 1) is set in a range from -1000 mV to 0mV.
Deviation max.2: 1000 mV
The maximum deviation between the two sensors (analog input 2 is higher than analog input 1) is set in a range from 0 mV to 1000mV.
Alarm Delay Time2: 9999 sec
An alarm will be released whenever the deviation between both inputs (analog input 2 and analog input 1) is lower/ higher thanconfigured Deviation min./max. values. The alarm can be delayed up to 9999 seconds in order to suppress short mismatches thatwould generate nuisance alarms.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 25
Equipment ConfigurationThe second screen in the Configuration Mode is:
Equipment Config Enter
Equipment Configuration Flow Chart
All of the parameters that define the specific operations of the equipment (dosing pumps) will be described in this section of theUser’s Manual.
Referring to the Equipment Configuration flow chart (Appendix A), use the keys on the control panel to respond to the informationon the LCD display (see System Flow Charts and Control Panel Keys).
The first screen in Equipment Configuration is:
Equipm. : 1 2 3 4Mode: R D 0 P
This screen displays the current configuration of each of the four dosing equipment. The different operational modes of the thisequipment have the following meanings:
R analog Regulator (Equipment 1 only) When this mode is configured the D-Determinator will dosechemical to maintain a desired setpoint as measured by thesensor configured on analog input 1.
F volume Fix (Equipment 1-3) In this mode the D-Determinator will dose in a fixed amountof chemical on a fixed time schedule.
C volume fix with secondary Control(Equipment 1 only)
In this mode the D-Determinator will dose in a fixed amountof chemical on a fixed time schedule and issue process alarmsfrom on the analog input connected to channel 1.
D Dependant (Equipment 2,3 only) In this mode the equipment connected to channels 2 and/or 3will dose it’s product in an amount that is calculate from aration of the amount of the main product that was dosed.
PorV
Pump or Valve on(Equipment 4 only)
This configuration is only for a larger pump connected tochannel 4 to provide a faster delivery of chemical into thesystem for initial charge and start-up mode.
0 Off (Equipment 2-4) No dosing equipment is configured on this channel.
Before the system has been configured for the first time, all pump configurations will be set to ‘0’. Press the ENTER key toproceed to the next screen.
Select EquipmentEquip. (1-4): 1
The cursor will be located in the digit field on the LCD display. Press the down arrow key, until ‘1’ appears in the digit field, toconfigure the first piece of equipment (dosing pump). Press the Enter key to proceed.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 26
Setting Equipment Mode (On/ Off)
Equipment Mode1: (I/O) = I
Every piece of equipment (dosing pump) may be activated or deactivated. The operation mode of each piece of equipment will beconfigured later. Rotate between the two choices, by using the down-arrow key.
If you do not want to use this pump on channel 1, you may Disable it by selecting ‘0’. When a pump is disabled, a ‘0’ will occupythat pump’s corresponding position in the ‘Equipment Mode’ display. To activate a pump connected to channel 1 then us thedown arrow keys to select a ‘1’.
When you have made your choice, press the ENTER key to confirm this screen.
Note: If a network of a Master and one or more Slaves are to be configured, care must be taken during configuration of theMaster:Every piece of equipment that is used by the network has to be configured at the Master, even if it is not used by theMaster. Only the Master has information about pump type and flow meter data.If an equipment is not used by the Master but used by one of the Slaves, act as follows:Set equipment to be on, define its data as necessary and to disable it at the Master set the "Operation Mode =Volume Fix (F)" and "Fixed Volume = 0,0 kg" or "Interval Time = 0 sec. With these settings, the equipment isdefined completely but is not dosing at the Master Control itself.
Select flowmeter
Prod. Flowmeter1: (Y/N) = Y
This option would be activated if there is a pulse flowmeters installed in the product dosing piping. The flow meter should beconnected to the corresponding flow meter input connection (i.e. for pump 1, it’s flow meter should be connected to Flow 1).
Flowmeters for equipment 2, 3 (i.e. pumps 2, 3) additional flowmeter may be activated or deactivated. The operation mode ofeach piece of equipment will be configured later. Rotate between the two choices (Yes/No), using the Down-arrow key and pressENTER when done.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 27
Select pump type
Select Pump Type1: (S/P) : S
For every piece of equipment (pump 1-4) the pump type can be selected between "Stroke" and “Permanent" (S/P). When you havemade your choice using the down-arrow key, press the ENTER key to confirm this screen.
If the pump type is selected to be "Stroke" (S), two additional times must be set:
Stroke Time1: 9,9 sec
"Stroke time" defines the length of time the stroke impulse is ON to drive the pump according to it’s specifications.
Pause Time1: 9,9 sec
"Pause Time" defines the pause time between two stroke pulses to produce the specified product flow rate (see parameter "StrokePump Product").
Flow Rate Input / Teach In
Prod.Flow.Rate1: T.I.(Y/N): N
Without use of a flow meter the discharge capacity (Mass Flow) of the pumps (kg/hr) can be set either by entering the flow rate ofthe pump (timed, l/hr or stroked, p/l) and the density of the product (kg/l), or by using a ‘Teach-in’ method. To choose betweenYes / No, toggle with the down arrow key. If a flow meter is used, its calibration is done simultaneously during 'Teach-in'. Ifthere is no flow meter and the Teach-In function is used then these are the screens that will be displayed. So, basically you dosethe chemical or water into a bucket using this feature to determine the flow rate of the dosing pump.
Fill Test Volume1: Start = Enter
Fill Test Volume1: Stop = Enter
Enter Weight1: 000,000 kg
Prod.Flow.Rate1: 250,0 kg/h
If a flow meter is used without the Teach-In calibration, "Flowmeter Product" has to be set (equipment 1-3) instead of the pumpparameters.
Pressing ENTER will start an internal timer and start the chemicaldosing per the current configuration of Stroke time and Pause time.
Pressing ENTER will stop an internal timer and stop the chemicaldosing.
Measure the amount of chemical dosed into the bucket in kilograms andenter it here. This screen will only be active for 1 minute so you mustenter the value in less then a minute or you have to run the test again.
The D-Determinator will now calculate a kg/h from the weightentered and the internal timer value.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 28
Product density will only appear with equipment 1-3 as equipment 4 uses the same product and flowmeteras equipment 1.
Stroke Pump Prod.1: 9999 p/l
OR
Flowmeter Prod.1: 9999 p/l
Product Density1: 1,999 kg/ l
Without a flowmeter one of the following parameters must be set, depending on selected pump type: "Timed Pump Prod" in caseof a permanent pump and "Stroke Pump Prod." in case of a stroke dosing pump. If equipment 4 is selected, these parametersappear always, as equipment 4 uses the same product and flowmeter as equipment 1.
Timed Pump Prod.1: 999,9 l/h
Stroke Pump Prod.1: 9999 p/l
Enter the pulses/l for the stroke pump if no flowmeter isconnected to the system.
Enter the pulses/l for the flowmeter connected to the system.
Enter the product density that will be dosed by the system.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 29
Control of product flowIf a flow meter is used then a minimum and maximum product flow rate can be set. During the dosing of product, flow ismeasured and compared with these limits. Whenever a flow outside these limits is measured, an alarm will be generated.
Min. Prod. Flow1: 9999,9 kg/h
Max. Prod. Flow1: 9999,9 kg/h
This parameter will be set for equipment 4 also if a flowmeter for equipment 1 is used since both pumps share the same flow meter.If equipment 4 has been selected, the configuration is finished now.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 30
Operation Mode
Select Operation ModeIf equipment 1 - 3 are selected, the operation modes can be set using these screens. The screen will display the configurationchoices of the selected equipment (i.e. equipment 1, 2,or 3). The configuration options are different forequipment 1 and equipment 2-3.
The choices for equipment 1 are the following:
Operation Mode1: (R/C/F/D) R
R Analog Regulator (This option is only available for equipment 1)C Volume fix with secondary Control (This option is only available for equipment 1)F Volume FixD Dependant (This option is only available for equipment 2 and 3)
For explanation of the different modes see the following chapters.
Note: The Operation Mode cannot be set for equipment 4, this equipment is used only for Initial Charge dosing of product1 and does not need any of the following settings. The Initial Charge volume is set during configuration ofequipment 1 as this equipment will be used for Initial Charge dosing if equipment 4 is disabled.
Analog RegulatorIn the analog regulator control mode the measured value from analog input 1 is compared to a desired setpoint. Depending on theselected Regulator Control Mode (discussed below) and the dosing operation configured on equipment 1 the D-Determinator willtry to maintain the entered setpoint:
Dosing of product 1 is done in portions defined by the parameters "Dose Time" and "Wait Time". If a dosing process is running, itwill be stopped after "Dose Time". Then the D-Determinator will wait until the end of the "Wait Time" to check the processconditions. If the dosing condition is still valid, dosing will be continued until the setpoint is reached or an alarm is generated.
Dose Time1: 9.999 sec
Wait Time1: 9.999 sec
Setpoint and Deadband are set in same units and scale as defined by the scale of analog input 1. For example if analog input 1 isdefined as conductivity with a range of 0....800 µS then the possible values for Setpoint and Deadband must be within that range,in µS.
Setpoint1: 0000
Deadband1: 0000
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 31
If the Setpoint cannot be reached within “Timeout Setpoint” time, an alarm is released and the automatic dosing process is stopped.If the measured value reaches the Setpoint during Initial Charge or Start-Up, these modes are stopped and a warning message isreleased.
Below is the screen to configure the Setpoint Timeout value.
Timeout-Setpoint1: 9999 sec
Reg. Crtl. Mode2: (Pos/Neg) Pos
Pos Positive Measured value should be above SetpointNeg Negative Measured value should be below Setpoint
Control Mode Positive:(Reg.Crtl. Mode = Pos)
• If the measured value is higher than "Setpoint" + "Deadband" dosing ofproduct 1 will be started.
• If the measured value is lower than "Setpoint" dosing of product 1 willbe stopped.
• If the measured value is below the alarm level "Min. Alarm" an alarm isgenerated and the automatic dosing process is stopped. If the measuredvalue is above the alarm level "Max. Alarm" a warning message isgenerated and the automatic dosing process continues.
Control Mode Negative:(Reg.Crtl. Mode =Neg)
• If the measured value is lower than "Setpoint" - "Deadband" dosing ofproduct 1 will be started.
• If the measured value is higher than "Setpoint" dosing of product 1 willbe stopped.
• If the measured value is below the alarm level "Min. Alarm" a warningmessage is generated and the automatic dosing process continues. Ifthe measured value is above the alarm level "Max. Alarm" an alarm isgenerated and the automatic dosing process is stopped.
As with Setpoint and Deadband the alarm levels are set in same units and scale as defined by the scale of analog input 1. Forexample if input 1 is defined as conductivity with a range of 0....800 µS then the possible range for Min./ Max. Alarm is the same.
Min. Alarm1: 0000
Max. Alarm1: 0000
The alarm warning level (Min. Alarm / Max. Alarm, depending on Regulator Control Mode) is not active during Initial Charge orStart-Up.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 32
Volume fixA Volume Fix mode control is open loop dosing of an amount of product per time, independent of any measured values. Twoparameters are necessary: volume and interval time. When an interval starts, a dosing process is started. If the set amount ofproduct has been dosed, the dosing is stopped until the end of the interval time. Then at the next interval time the dosing is startedagain.
Fixed Volume1: 999,999 kg
Interval Time1: 9.999 sec
The consumption of product is controlled via flow meter if possible else the calculation of the consumption is done via time andthe calculated product flow rates of the equipment.
Volume fix + secondary ControlA Volume Fix mode is dosing an amount of product per time independent of any measured values. Two parameters are necessary:volume and interval time. When an interval starts, a dosing process is started. If the set amount of product has been dosed, thedosing is stopped until the end of the interval time. Then the next interval is started again. As a supporting feature, the measuredvalue is used to control the action of D-Determinator. If the measured value exceeds the alarm limits then alarms are generated,but the dosing and control is still maintained by the volume fix dosing operation.
Fixed Volume1: 999,999 kg
Interval Time1: 9999 sec
The consumption of product is controlled via flow meter if possible else the calculation of the consumption is done via time andthe set product flow rates of the equipment.
The alarm levels are set in same units and scale as defined by the scale of analog input 1. For example if analog input 1 is definedas conductivity with a range of 0....800 µS then the possible range for Min./ Max. Alarm is the same.
Min. Alarm1: ?????? ???
Max. Alarm1: ?????? ???
The alarm warning level (Min. Alarm / Max. Alarm, depending on Regulator Control Mode) is not active during Initial Charge orStart-Up.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 33
DependantIn dependant mode, the amount of product 2 or 3 is set in relationship to product 1. If, for example, a value of 0..200 is entered forthe relationship, the equipment will dose, on average, 20% of the mass of this product compared to the mass of product 1.
To prevent the equipment being started too often, a minimum volume is defined. If the need for product 2 or 3 exceeds this limit, adosing process is started dosing the complete necessary amount of product.
min. Volume2: 999,999 kg
Vol. Ratio2: 1,000
The consumption numbers are used to determine when it time to dose, not an actual ratio flow that runs in parallel with product 1,just at a lower flow rate.
The consumption of product is controlled via flow meter if possible else the calculation of the consumption is done via time andthe set product flow rates of the equipment.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 34
Setpoints for Initial Charge and Start-UpFor every piece equipment the amount of product necessary for an initial charge has to be set here. During Initial Charge andStart-Up the measured value in analog input 1 is monitored.
Setp. Init. Ch.1: 999,999 kg
Setp. StartUp1: 999,999 kg
The dosing of the configured Initial Charge/Start-up volumes can be manually activated within the Operation Menu or externallytriggered through the hardware inputs for Initial Charge trigger and Start-up trigger (see Control System Configuration).
Return to Configuration Menu
Return to ConfigStart = Enter
To exit Equipment Configuration, and return to the Configuration Menu, press the ENTER key at this screen.
Reading / Resetting Total ConsumptionWithin the Configuration menu press SELECT to get into the part of the menu, until this screen is displayed allowing to reset totalconsumption counters.
Reset Cons. Enter
After the system has displayed the total consumption amount for equipment 1-3, you may reset the total product consumptionamount for each equipment to zero, after entry of the password.
Consumption/tot.1: 99999,9 kg
Consumption/tot.2: 99999,9 kg
Consumption/tot.3: 99999,9 kg
Although the total consumption is displayed separately for each equipment, resetting will be done simultaneously for all threetotals on all three products.
Consumption/totReset = Enter
Enter Password Xxxx
If a false password is entered, the "Reset Consumption" screen will be shown again.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 35
Network setting
If it the application requires a CAN-Bus connection between the D-Determinator units. An additional CAN-Bus interface moduleneeds to be purchased along with the D-Determinator to complete a network installation.
If the D-Determinator is equipped with a CAN-Bus interface, this option appears in the menu.
Network setting Enter
The type of operation in the network has to be selected first.
Use Network(M/S/N): M
Three selections are available:M Master This D-Determinator is a Master to other units within the same network.S Slave This D-Determinator is Slave to another units within the same networkN None The CAN-Bus interface will not be used in conjunction with other DiverseyLever Controls
Network ID32
The address within the CAN-Bus network has to be set.
If the D-Determinator is selected to be a "Master" then the address has to be a multiple of 4. A Master will regard all D-Determinators as slave if their address is higher by 1…3 compared to the Master’s address.
Note: If another address is set, it will be reduced to the next lowest multiple of four.Example: If the Master address is set to be 19 it will be changed automatically to 16 if ENTER is pressed.
If a D-Determinator is selected to be a "Slave" it’s address must not be a multiple of four. This slave will regard a D-Determinatoras Master if his address is the next lowest multiple of four compared with his own address. Care must be taken not to assign twoslaves to the same address.
Note: If a multiple of four is set, the address will be incremented by one.Example: If the slave address is set to be 20 it will be changed automatically to 21 if ENTER is pressed.
Network Example: If the Master D-Determinator’s CAN-Bus Address is 32 then it’s Slave Addresses will be 33, 34, 35
Changing the Password
Change Password Enter
The four-digit password can be changed when a new code is entered twice in a row. When the two entries don’t match, the newpassword will not be accepted.
Repeat Password xxxx
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 36
After entering a false code the system switches back to entry of the password.
Note: System is delivered with password 9999, which can be modified!
Input / Output Test
I/O - Test Enter
The reliability of the process I/O (analog input, pulse flow, digital input and digital output channels) of theD-Determinator system can be verified using these screens (refer to Appendix A for flow chart).
Analog Input1: 06,44 mA
Analog Input2: 08,28 mA
These two screens are used to verify that the analog input signals from the two field sensors if correct. The values displayed willbe the current values on the current loop. Typically, the commissioning personnel should check the field signals per this table.
Sensor Value 0-20mA Configuration 0-20mA ConfigurationValue that should be displayed Value that should be displayed
Zero value 0,00 4,00Middle value of the sensorrange
10,00 12,00
Full Value of the sensor range 20,00 20,00
Flow sensor 1-8x x x x x P P P
The status of the flow sensors will be displayed next with ‘1’ indicating a ‘ON’ or closed contact, 'P' indicating pulses at the inputand ‘0’ indicating a ‘OFF’ or open contact. From left to right the indications are for flow sensor 1, flow sensor 2 and flow sensor 3
Dig. Input 1-8x x 0 1 0 1 0 0
The status of most digital inputs will be displayed, with ‘1’ indicating a ‘high’ or ON, and ‘0’ indicating a ‘low’ or OFF signallevel. The Digital Input from left to right are:
Low Level Switch Equipment 1 (Contacts - LA1 & GND)Low Level Switch Equipment 2 (Contacts - LA2 & GND)Low Level Switch Equipment 3 (Contacts - LA3 & GND)Start-up external request (Contacts - Start & GND)Initial Charge external request (Contacts - Init & GND)External Trigger (Contacts - Trig & GND)Rack Counter (Contacts – Rack1 & GND) {Note: This signal is not on the Input Test screen}
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 37
Set Output 1-80 0 0 1 0 1 0 0
Outputs from left to right. To toggle the output press the down arrow key and to move to the next output press side arrow key.
Stroke Pump 1 (Contacts - S1C & S1C)Stroke Pump 2 (Contacts - S2C & S2C)Stroke Pump 3 (Contacts - S3C & S3C)Stroke Pump 4 (Contacts - S4C & S4C)Permanent Pump 1 (Contacts - P1a & P1a and P1b & P1b)Permanent Pump 2 (Contacts - P2a & P2a and P2b & P2b)Permanent Pump 3 (Contacts - P3a & P3a and P3b & P3b)Permanent Pump 4 (Contacts - P4a & P4a and P4b & P4b)
Set Output 9-16x x x x x x 0 0
Alarm Contact (Contacts - Alarm and Alarm)Run Contact (Contacts - Run and Run)
All digital outputs for the 4 pumps, 4 stroke signals to the pumps and the output for the alarm and the running message can beindividually set or reset by toggling between ‘1’ and ‘0’ using the down arrow key. After exiting the Input / Output Test, allprevious output signal levels will be restored.
The order of the signals is the first four outputs are the Stroke PumpOutputs (1, 2, 3, & 4). The next four are the Permanent Pump Outputs (1, 2, 3, & 4).
The order of the signals is the first channel is the Alarm relay and thesecond channel is the Run relay.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 38
Resume Operation
Resume OperationStart = Enter
To exit the Configuration Mode and resume operation in the Automatic Mode, press the Enter key. Immediately upon exiting theConfiguration Mode, the system will begin sequencing in Automatic Mode, using the configuration parameters, that were lastentered.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 39
Operating instructions
Operation of D-Determinator
The D-Determinator starts to operate automatically after power up (On/Off switch on the lower right side of the unit). Powersupply is indicated by an activated LC display.
Conduct. No Rel.0234 mS 23,5°C
(In the screen example above: Conductivity sensor is connected to analog input 1, Temperature sensor is connected to analog input2, D-Determinator waiting for a release trigger)
The display always shows the configured input type, the operation mode and the values at both analog inputs.
If the display shows "Automatic" the main dosing process is running. A external release trigger signal is necessary for automaticoperation.
RedOx Autom.0234 mV 35,0°C
(In the screen example above: RedOx-sensor is connected to analog input 1, Temperature sensor is connected to analog input 2, D-Determinator is dosing automatically per the last entered configuration). The display may be switched at any time to display of thesetpoint by pressing the "⇒" key. The operating display is restored as soon as the key is released.
Setpoint0250 mV
If the display shows "Initial Charge", "Start-Up" or "Manual", an additional dosing process is running. To start an Initial Chargeor a Start-Up dosing an additional field input signal (Initial Charge or Start-Up) is required. These modes may be triggeredmanually too (see below). The configuration can be changed at any time (refer to "Programming the D-Determinator").
pH Init.Ch.08,8 pH 48,6°C
(In the screen example above: pH-sensor is connected to analog input 1, Temperature sensor is connected to analog input 2, D-Determinator is dosing an Initial Charge)
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 40
ClO2 Start-Up125,4 ppm 61,2°C
(In the screen example above: ClO2 sensor is connected to analog input 1, Temperature sensor is connected to analog input 2, D-Determinator dosing in Start-Up mode)
Manual dosing can only be started through the LCD screen and an operator request using the SELECT key andENTER key (see below).
Manual Equip.108,8 % 35,0°C
(In the screen example above: Conductivity sensor is connected to analog input 1 displaying in % of a concentration, Temperaturesensor is connected to analog input 2, D-Determinator is dosing by equipment 1, manually). The display may be switched at anytime to display of the setpoint and the remaining volume by pressing the "⇒"-key. The display is restored as soon as the key isreleased.
Setpoint Volume150,0 ppm 065,9kg
(In the screen example above: ClO2 sensor is connected to analog input 1, D-Determinator is dosing in Initial Charge/ Start-Up/Manual Mode with a remaining volume of 65,9 kg is to be dosed)
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 41
Automatic dosingDuring automatic dosing the operation depends on the selected operation mode, see chapter "Equipment Configuration", and onthe configuration of the equipment.
Once configured Equipment 1 (main product) can dose product with these control algorithms, "Analog Regulator", "Volume Fix +2nd Control" or "Volume Fix", see below for details of these modes. Equipment 2 and 3 can use, simultaneously to Equipment 1,these control algorithms "Volume Fix" or "Dependant" mode.
Analog RegulatorThe measured value from analog input 1 is compared to a setpoint. Depending on the selected Regulator Control Mode thefollowing happens:
Control Mode Positive:(Reg.Crtl. Mode = Pos)
• If the measured value on analog input 1is higher than "Setpoint" +"Deadband" dosing of product 1 will be started.
• If the measured value on analog input 1is lower than "Setpoint" dosingof product 1 will be stopped.
• If the measured value on analog input 1is below the alarm level "Min.Alarm" an alarm is generated and the automatic dosing process isstopped. If the measured value on analog input 1is above the alarmlevel "Max. Alarm" a warning message is generated and the automaticdosing process continues.
Control Mode Negative:(Reg.Crtl. Mode = Neg)
• If the measured value on analog input 1is lower than "Setpoint" -"Deadband" dosing of product 1 will be started.
• If the measured value on analog input 1is higher than "Setpoint" dosingof product 1 will be stopped.
• If the measured value on analog input 1is below the alarm level "Min.Alarm" a warning message is generated and the automatic dosingprocess continues. If the measured value on analog input 1is above thealarm level "Max. Alarm" an alarm is generated and the automaticdosing process is stopped.
Dosing of product 1 is done in portions defined by the parameters "Dose Time" and "Wait Time". If a dosing process is running, itwill be stopped after "Dose Time". D-Determinator will wait until the end of "Wait Time". If the dosing condition is still valid(see table above for dosing conditions), dosing will continue.
The consumption of product is monitored via flowmeters if possible else the calculation of the consumption is done via time andthe calculated product flow rates of the equipment. The consumption counters are updated each time the product is dosed.
If the Setpoint cannot be reached within the dosing time, an alarm is generated and the automatic dosing process is stopped untilthe process operator acknowledges the alarm by hitting the ENTER key.
Volume fix + secondary ControlA Volume Fix mode is dosing an amount of product per time independent of any measured values. Two parameters are necessary:volume and interval time. When an interval starts, the main dosing process is started. If the set amount of product has been dosedwithin the interval time, the dosing is stopped until the end of the interval time. Then at the next interval the dosing process isstarted, again. As a supporting feature, the measured value is used to control the action of D-Determinator.
The consumption of product is monitored via flowmeters if possible else the calculation of the consumption is done via time andthe calculated product flow rates of the equipment. The consumption counters are updated each time the product is dosed.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 42
As a supporting feature, the measured value is used to control the action of D-Determinator. Depending on the selected RegulatorControl mode the following happens:
Reg.Crtl. Mode = Pos.: If the measured value on analog input 1is below the alarm level "Min. Alarm" an alarm is released andthe automatic dosing process is stopped. If the measured value is above the alarm level "Max. Alarm" awarning message is released and the automatic dosing process continues. The warning message willremain on the screen until the process operator acknowledges it by hitting the ENTER key.
Reg.Crtl. Mode =neg.: If the measured value on analog input 1is below the alarm level "Min. Alarm" a warning message isreleased and the automatic dosing process continues. If the measured value is above the alarm level"Max. Alarm." an alarm is released and the automatic dosing process is stopped. The warning messagewill remain on the screen until the process operator acknowledges it by hitting the ENTER key.
Volume fixA Volume Fix mode is dosing an amount of product per time independent of any measured values. Two parameters are necessary:volume and interval time. When an interval starts, a dosing process is started. If the set amount of product has been dosed, thedosing is stopped until the end of the interval time. Then during the next interval the dosing process is started, again.
The consumption of product is controlled via flowmeter if possible else the calculation of the consumption is done via time and theset product flow rates of the equipment.
DependantIn dependant mode, the amount of product 2 or 3 is set in relationship to product 1. If, for example, a value of 0,200 is entered forthe relationship, the equipment will dose, on average, 20% of the mass of his product compared to the mass of product 1. Thisamount is based on the consumption values not on simultaneous ratio feeding.
To prevent the equipment being started too often, a minimum volume is defined. If the need for product 2/3 exceeds this limit, adosing process is started dosing the complete necessary amount of product.
The consumption of product is monitored via flowmeters if possible; if not the calculation of the consumption is done via time andthe calculated product flow rate of the equipment.
Dosing an Initial Charge/ Start-UpDuring Initial Charge/ Start-Up a preset volume is dosed, it is possible to dose the initial charge or the start-up charge with alldosing equipment (1-4). The equipment will run at maximum flowrate to perform the initial charge or the start-up charge asquickly as possible. Equipment 4 is a special case because it will dose the main product with a larger pump if configured.
To start an Initial Charge or a Start-Up dosing, externally, two input signals (Release Trigger plus Initial Charge or ReleaseTrigger plus Start-Up) are necessary. Once an Initial Charge or Start-Up is started externally the D-Determinator will dose theentire amount configured unless the Release Trigger signal is lost.
If an Initial Charge or a Start-Up dosing is running, the one of following screens will appear in the menu.
Conduct. Init.Ch0555 mS
Pressing the SELECT-key will lead to a screen allowing to stop the Initial Charge/ Start-Up Mode manually. If it was triggeredexternally.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 43
Initial ChargeStop = Select
Start-Up ModeStop = Select
Initial Charge ENTER
Start-Up Mode ENTER
These modes may be started manually, also. If the Release Trigger is active, both following screens will appear in the menu. Theoperator can loop through the menu by hitting the SELECT key.
Initial Charge ENTER
Start-Up Mode ENTER
Pressing the ENTER-key will lead to a screen allowing to start the Initial Charge/ Start-Up Mode manually
Initial ChargeStart = Enter
Start-Up ModeStart = Enter
The volumes to be dosed and the selected equipment have to be set in Configuration Mode. Each piece of equipment (1-4) thathave an initial charge or start-up amount configured will dose that product when one of these modes are selected, either through thescreen and operator selection or by an external trigger.
In these modes the warning level is deactivated (either Min. Alarm" or "Max. Alarm" depending on Control Mode) only the Alarmlevel will be active (see chapter "Equipment Configuration").
During Initial Charge, equipment 4 is used instead of equipment 1 (if this equipment is configured to be "ON").
If the Operation Mode is "Analog Regulator" these modes are stopped once the setpoint is reached, D-Determinator will continuein automatic mode. In this case a warning message will be released.
If the Operation Mode is "Volume fix + 2nd Control" these modes are stopped once the alarm level is reached. In this case analarm message will be released and the process will be stopped.
If the Operation Mode is "Volume fix " no alarm level are activate.
Initial Charge/ Start-Up Mode are finished once the defined volumes of all products have been dosed.
Manual dosingDuring Manual Mode a preset volume is dosed, it is possible to choose between Equipment 1, 2 and 3. The equipment will run atmaximum flowrate to deliver the desired amount of product as quickly as possible. The dosed volume is set while starting ManualMode operation.
ENTER
ENTERThese second screens can beleft without stopping InitialCharge / Start-Up by pressingthe ENTER-key
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 44
In this mode the warning levels are deactivated (either Min. Alarm" or "Max. Alarm" depending on Regulator Control Mode) onlythe Alarm level will be active (see chapter "Equipment Configuration").
the following screen will appear in the menu after viewing the consumption:
Manual Mode ENTER
Pressing the ENTER-key will lead to operator to screens allowing them to enter the necessary information to complete the desiredManual dose operation:
The equipment number and the desired product volume have to be selected and entered, respectively:
Manual ModeEquipm. (1..3):2
Manual Volume2: 999,9kg
The Manual Mode has to be started by pressing the ENTER-key.
Manual Mode2: Start = Enter
While dosing manually, the remaining time can be displayed by pressing the "⇒"-key during normal display. If the Manual Modeis already running, a different screen will appear when selecting this mode.It can be stopped then by pressing the SELECT-key
Manual Mode2: Stop = Select
Pressing "ENTER" continues dosing and returns to normal operation.Pressing "SELECT" stops the manual dosing and returns to normal operation.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 45
Operation as a "Slave"If a D-Determinator is used as Slave the following differences in operation will occur:
− The sensor input 1 is of the same type as defined at the Master.
− The equipment will always be "permanent" as valves will be controlled and not dosing pumps.
− The valves 1-4 corresponding to equipment 1-4 have to be connected to the power outputs, the stroke outputs will be inactive.
− Level alarm, data of equipment and flow sensor inputs will not be used, these are only necessary for the Master. If the networkaddress of a D-Determinator is set to a Slave-Address, these items will be skipped during configuration.
− During Initial Charge output 4 (equipment 4 = valve 4) will be activated. This allows to use a separate valve for dosing theInitial Charge. If the same valve (valve 1) is used to control the flow of product 1 during normal operation and during InitialCharge, output 4 has to connected to output 1
Network operationIf multiple D-Determinators are coupled together using the CAN-Bus network some rules of operation are necessary:
− Every slave will send a request of product to its Master if product is needed. The Master will do the same (internally).
− The unit to be serviced first is the unit that sends the first request (FIFO - principle: first in - first out).
− During Wait Time (Regulator Mode) another unit may be serviced.
− An Initial Charge or a Start-Up dosing may not be interrupted by another unit.
−
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 46
Troubleshooting
Alarms and WarningsThere are two types of messages: Alarms and Warnings.
Alarm: If a condition in the process triggers an alarm the dosing process will be stopped until the alarm is acknowledged bypressing the ENTER-key.
Warning: If a warning message is generated the D-Determinator continues with normal operation. The warning message willstay on the display until it is acknowledged by pressing the ENTER-key
An alarm / warning will always be signaled in the top line of the display. If possible, the bottom line will continue to show themeasured values of both analog inputs.
The last alarm or warning that occurred will be displayed. If an alarm is activated, no warning message will override this display.
Alarm messsages
E1 min. Deviation????? ??? ???? ??
E2 max. Deviation????? ??? ???? ??
The difference between the measured value at both inputs are higher then the set limits. This is possible if input 1 is RedOx orClO2 and input 2 is RedOx, using secondary control for input 1
E3 broken wire 1- - - - - ??? ???? ??
E3 broken wire 2- - - - - ??? ???? ??
The input current on analog input 1 or 2 is lower than 2 mA. This is only possible if the input has been defined as a4-20mA input.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 47
E4 Config. 2????? ??? - - - - °C
Input 2 has not been configured although it is essential for the dosing operation. This is possible if input 1 is a conductivity sensorwithout temperature compensation and input 2 has not been configured as temperature input.
E5 XXXX min.????? ??? ???? ??
The measured value is below the low Alarm threshold (Regulator Control Mode = positive). "XXXX" denotes the selected inputtype, e.g. "RedOx" or "Cond."
E5 XXXX max.????? ??? ???? ??
The measured value is above the high Alarm threshold (Regulator Control Mode = negative). "XXXX" denotes the selected inputtype, e.g. "RedOx" or "Cond."
Setpt.nt reached????? ??? ???? ??
The setpoint could not be reached within the set timeout limit (Analog regulator Mode, equipment 1 only).
Vol.nt reached X????? ??? ???? ??
The preset volume could not be dosed within the set interval time. (both Volume fixed Modes).The demand for volume is higher then ten times the preset minimum volume (Dependant Mode)."X" denotes the equipment (X=1, 2 or 3) generating this error
E7 Level Alarm X????? ??? ???? ??
The low level alarm input from product "X" has been activated (X=1, 2, 3). An alarm is generated if the low level alarm input isactivated (the contact is open) indicating that the related equipment/product is empty and the process is requesting product.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 48
Warning messages
W1 Low Flow Pr.X????? ??? ???? ??
The flow of product "X" (X=1, 2, 3) has been measured and is below the set minimum flowrate.
W2 High Flow Pr.X????? ??? ???? ??
The flow of product "X" (X=1, 2, 3) has been measured and is above the set maximum flowrate,.
W5 XXXX min.????? ??? ???? ??
The measured value is below the low warning threshold (Regulator Control Mode = negative). "XXXX" denotes the selectedinput type, e.g. "RedOx" or "Cond.".
W5 XXXX max.????? ??? ???? ??
The measured value is above the high warning threshold (Regulator Control Mode = positive) ). "XXXX" denotes the selectedinput type, e.g. "RedOx" or "Cond.".
W7 Level Alarm X????? ??? ???? ??
The low level alarm input of product "X" has been activated (X=1, 2, 3). A warning is generated if the low level alarm input isactivated, indicating the related equipment/ product is empty and the process is not requesting product. This error exists aswarning and alarm. As longs as a product is not demanded, no alarm is generated because a Master not needing this productshould not be stopped by a level alarm of this product. In case of a product rarely needed, the equipment may continue working aslong as no actual demand appears.
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 49
System Not FunctionalIf no display appears on the LCD screen when the system is switched on:
• check all electrical connections
• ensure that all fuses are functioning and not broken
• ensure that the power supply has not been disrupted
If the system still does not function, please contact the service department.
If the display is working, check inputs and the reaction of the outputs with the integrated "I/O-Test" function
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 50
Example
In this example configuration the D-Determinator is configured to Analog Regulator dose product 1 and Dependant Dose product2 and product 1 has a flow meter installed.
Configuration Parameter Configured ValueEnter Password 9999Sel. Language SELECT KeySensor Config. ENTER KeySelect Sensor 1: Conduct.Sel. Current Min. 1: (0/4mA) 4Cond.: Sel. Unit 1: mSCond.: Scale max 1: 1500Cond.:Temp.Comp 1: (Y/N) YSelect Sensor SELECT KeySelect Sensor 2: Temp..Sel. Current Min. 2: (0/4mA) 4Cond.: Scale max 2: 80Select Sensor SELECT KeySensor Config. SELECT Key
Equipment Config. ENTER KeyEquipment Display Equip. 1 2 3 4
Mode: R D 0 0Equipment Select. Equip. (1-4) : 1Equipment Mode 1: (I/O) 1Prod. Flowmeter 1: (Y/N) YSelect Pump Type 1: (S/P) PProd.Flow.Rate 1: T.I. (Y/N) N This function is used to calibrate the
dosing pump. If you need to calibratethe pump select Y and follow thescreen instructions
Flowmeter Prod. 1: 20 p/lProduct Density 1: 1,100 kg/lMin. Prod. Flow 1: 2 kg/hMax. Prod. Flow 1: 100 kg/hOperation Mode 1: (R/C/F): RDose Time 1: 5 secsWait Time 1: 10 secsSetpoint 1: 0800 msDeadband 1: 0100 msTimeout-Setpoint 1: 0020 secReg. Cnt. Mode 1: (Pos/Neg) NegMin. Alarm 1: 0500 mSMax. Alarm 1: 1000 mSSetp. Init.Ch 1: 001,000 kgSetp. Startup 1: 000,500 kgEquipment Select. Equip. (1-4) : 2Equipment Mode 2: (I/O) 1Prod. Flowmeter 2: (Y/N) NSelect Pump Type 2: (S/P) SStroke Time 2: 5 secsPause Time 2: 9 secs
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 51
Prod.Flow.Rate 2: T.I. (Y/N) N This function is used to calibrate thedosing pump. If you need to calibratethe pump select Y and follow thescreen instructions
Stroke Pump Prod. 2: 1000 p/lProduct Density 2: 1,000 kg/lOperation Mode 2: (D/F): DMin. Volume 2: 001,000 kgVol. Ratio 2: 0,100Setp. Init.Ch 2: 000,010 kgSetp. Startup 2: 000,005 kg
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 52
Appendix A: Specifications
Supply Voltage : 115 V, 50...60 Hz or230 V, 50...60 Hz (selectable by jumper J1)
Output to equipment (permanent) : Relays with two contacts,Contacts may be connected to 24V/DC or supply voltage
Max. output current (permanent) : 3 A (8A peak) per contact
Output to equipment (stroke) : Optocoupler, may be connected to 24V DCMax. output current : 0,1 A
Output voltage ("Alarm, Running"): Relays with one contact,Contact may be connected to 24V/DC or supply voltage
Max. output current (permanent) : 3 A (8A peak) per contact
Output voltage to flow sensor : 24V DC stabilizedMax. output current : 30 mAInput voltage : 5V.. 24V
Output voltage to analog sensor : 24V DC, stabilizedMax. output current : 30 mA
Inputs (ext. Trigger, Init. Charge : internal pull-up resistor (6,8 kOhm) to +24V DC Level Alarm)
Internal voltages : 24 V, DC (accessible): 5 V, DC (accessible)
Timing range : 1 ... 9999 secondsTiming tolerance : 0,1 %
Operator interface : Integrated LC-display (2 lines with 16 characters): 4 keys
Connectors : spring loaded terminals, max. 1,5mm2
Weight : 2,5 kgSize : L 255 mm x W 323 mm x H 85 mmBox : ABSInsulation protection : IP 65Temperature range : -10°C ... +60°C
Transformer
Spring Loaded TerminalsOn
J3
Off
115
J1
230
Fuse6,3 AT
Fuse1 AT
SpanInput 1
SpanInput 2
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 53
Appendix B: Electrical Connections
The D-Determinator system should be connected, as the diagram below:
AI AI PI PI PI DI DI DI DI DI DI DI DO DO DO DO DO DOC
AN
-Bus
CA
N-B
us
Ana
log
Inpu
t 1
Ana
log
Inpu
t 2
Inpu
t Flo
w S
enso
r 1 (P
ulse
)
Inpu
t Flo
w S
enso
r 2 (P
ulse
)
Inpu
t Flo
w S
enso
r 3 (P
ulse
)
Leve
l Ala
rm 1
Leve
l Ala
rm 2
Leve
l Ala
rm 3
Sta
rtup
Trig
ger
Initi
al C
harg
e T
rigge
r
Run
Trig
ger
Rac
k In
put (
Mac
hine
Spe
ed)
Ala
rm R
elay
Run
Rel
ay
Pum
p 1
Str
oke
Rel
ay
Pum
p 2
Str
oke
Rel
ay
Pum
p 3
Str
oke
Rel
ay
Pum
p 4
Str
oke
Rel
ay
CH CH + + + + + Alm1 Run S1C S2C S3C S4CPE PE AI1 AI2 FI1 FI2 FI3 LA1 LA2 LA3 Start Init Trig Rack1 PE PE PE PE PE PECL CL ⊥ ⊥ ⊥ ⊥ ⊥ GND GND GND GND GND GND GND Alm1 Run S1E S2E S3E S4E
PWR PWR PWR PWR DO DO DO DO DO DO DO DO PWR PWR PWR PWR
Ava
ilabl
e 24
VD
C P
ower
Ava
ilabl
e 24
VD
C P
ower
Ava
ilabl
e 24
VD
C P
ower
Ava
ilabl
e 24
VD
C P
ower
Pum
p 1
DC
con
trol
Pum
p 1
AC
con
trol
Pum
p 2
DC
con
trol
Pum
p 2
AC
con
trol
Pum
p 3
DC
con
trol
Pum
p 3
AC
con
trol
Pum
p 4
DC
con
trol
Pum
p 4
AC
con
trol
Sup
ply
Vol
tage
Pow
er (
115V
AC
or
230V
AC
)
Sup
ply
Vol
tage
Pow
er (
115V
AC
or
230V
AC
)
Sup
ply
Vol
tage
Pow
er (
115V
AC
or
230V
AC
)
Sup
ply
Vol
tage
Pow
er (
115V
AC
or
230V
AC
)
Det
erm
inat
or In
put P
ower
+ + + + P1A P1B P2A P2B P3A P3B P4A P4B L1 L1 L1 L1 L1PE PE PE PE PE PE PE PE PE PE PE PE PE PE PE PE PE
GND GND GND GND P1A P1B P2A P2B P3A P3B P4A P4B N N N N N
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 54
Appendix C: Menu Schemes
Operation MenuThe two control panel keys, Select and Enter, are indicated on the flow charts by ‘S’ and ‘E’, respectively.
From configurationfrom any screen when no key was pressed for 60 seconds
Status or Alarmdisplay
View Consumption Enter
Select EquipmentEquip. (1-3): 2
Consumption/tot.2: 99999,9 kg
Consumption/day2: 999,999 kg
Manual Mode Enter
Manual ModeEquipm. (1..3): 2
Manual Vol.=123,42:Stop = Select
when Manual Mode is complete
Initial charge Enter
Configuration Enter
Enter Password xxxx
correct password
Incorrect password
S
S
S
E E
E E E
S/E
ð
S
S
E E E E
System start when not configured
Go to configuration
Consumption
Manual Mode
if Initial chargeis active
Consumption/day2: Reset = Enter
Do resetconsumption/day
E E
Go Offline
Initial ChargeStop = Select
E
Initial Charge
only if Initial Charge is activated
S
Manual Volume2: 999,9 kg
Manual Mode2: Start = Enter
E
Loop ThruEquip.No. 1-3
S
S S
SS
Stop Initial Charge
Start Up Mode Enter
if Start Upis active
Start Up ModeStop = Select
E
Start Up Mode
only if Start Up Mode is activated
SStop Start Up Mode
Status and Alarm display(types see table below)
only if Initial Charge is not activated
only if Start Up Mode is notactivated
Start Up ModeStart = Enter
E EStart Start Up Mode
Initial ChargeStart = Enter
E E Start Initial Charge
if an alarm msg is displayed: E to acknowledge the alarm
IfAut
Ye
No
Setpoint1000 mS ò E+1 E+2 E+3 E+4
F:0 D:1 0:0 0:0The D-Determinator will display theStatus or Alarm display until anoperator key is selected.
ò
S
ConfigurationOperations
S
when Manual Mode is already active
S
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 55
Configuration MenuThe two control panel keys, Select and Enter, are indicated on the flow charts by ‘S’ and ‘E’, respectively.
To Operationfrom any screen whenno key was pressed for 60s
Reset Cons. Enter
Consumption/tot.1: 9999,9 kg
Change Password Enter
I/O-Test Enter
S
S
S
E E E
From Operation menu
Reset Consumption
Enter Password xxxx
Repeat Password xxxx
Do Change PasswordE E E
Change Password
S S
Analog Input 1 06,44 mA
Set Output 1 - 8 0 0 1 0 1 1 1 1
E E
I/O-Test
S
Resume OperationStart = Enter
S
Consumption/ tot.2: 9999,9 kg
Enter Password xxxx
Do Reset Consumption/totfor product 1, 2, 3
E
E
Consumption/ tot.3: 9999,9 kg
Set Output 9 -16x x x x x x 0 0
E
ES
Flow Sensor 1-8x x x x x P P P
ES
Analog Input 2 03,63 mA
Dig.Input 1-8x x 0 0 1 1 0 0
E
S S S
S
Consumption/tot.Reset = Enter
E
S
Network Setting Enter
S
Use Network(M/S/N): M
Network ID1234
E E E
Network Settings
S S/ false Code
M: Master to other DD-SlavesS: Slave to DD-MasterN: No usage of Network
Equipment Config Enter
S
Sensor Config. Enter
S
Sensor ConfigurationE
Equipment ConfigurationE
Goto Automatic Mode(Operation menu)
(false code)
(false code)
Sel. Language Enter
Sel. Language(E/D/F/Sp): E
D-DeterminatorVersion 1.21
E E S/E
Language Settings
S
S
Enter Password 9999
IfCAN-
Yes
No
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 56
Input Configuration, Submenu to ConfigurationThe two control panel keys, Select and Enter, are indicated on the flow charts by ‘S’ and ‘E’, respectively.
Select Sensor1: Conduct. pH-Scale min.
1: 00,0pH-Scale max.1: 14,0
Calibrate Sensor1: (Y/N) Y
Calibrate pH=71: 07,1 [ENTER]
Calibrate pH=4/101: 09,8 [ENTER]
E
S E
E EE
No
Yes
E E
From Configuration
ConductivitySensor
pH-Sensor
RedOx-Sensor
Cond.-Scale max.1: 2000 mS
Cond.:Temp.Comp. 1: Y/N
Cond. sel. Unit1: mS
Cond.:Temp.Coeff 1: 10,0 %/°C
Cond.:Ref. Temp. 1: 25,0 °C
E
E
Yes
E E
NoFree Cond. Curve1-1:0500mS=07,5%
only if "curve"is selected
RedOx-Scale min.1: -1000 mV
RedOx-Scale max.1: 1000 mV
E E E
ClO2-Sensor
ClO2-Scale max.1: 1000 ppm
E E
Select Sensor2: Temp.
Direct display of sensor signal
Direct Display1: (mA/%): mA
E E
"RedOx"
"pH"
"Conduct."
"ClO2"
"direct"
RedOx-Sensor (only if Sensor 1 is RedOx or ClO2)
E
Temperature -Sensor
Temp.-Scale max.2: 080 °C
E E
"Temp."
To Configuration
Secondary Contr.2: (Y/N) Y
Yes
No
Deviation min.2: -10 %
E Deviation max.2: +10 %
E Alarm delay time75 sec
E
no auxiliary Sensor
deactivate Aux. InputE
"RedOx"
"off"
This option is chosen automatically if"Main sensor" = "Cond."and "Temp.-Compens."="active"
Slave-Controluses thesamesensor asit’s Master
in case of two RedOx-Sensor,the scale is the same for both sensors
Sel. Current Min.1: (0/4mA): 4
Sel. Current Min.2: (0/4mA): 4
S E S
S
E E
(All Sensortypes branchto this screen)
(Range 1 – 2000)(Cond. Units **)
** Cond. Units are (mS, uS, NaOH, KOH, H3PO4, H2SO4, HNO3 or curve)
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 57
Equipment Configuration, Submenu to ConfigurationThe two control panel keys, Select and Enter, are indicated on the flow charts by ‘S’ and ‘E’, respectively.
Select equipment(1-4): 1
S =Stroke
S
From Configuration
switch off equipmentE “0"
To Configuration
Product Density1: 1,999 kg/l
Prod. Flow Rate1: T. I. (Y/N):Y
Fill Test Volume1: Start=Enter
Enter Weight1: 999,9 kg
Fill Test Volume1: Stop =Enter
Yes
No
Setpoint1: ????? ???
WaitTime1: 999 sec
Timeout - Setpoint 1: 999 sec
Deadband1: ????? ???
Interval Time1: 999 sec
fixed Volume1: 999,9 kg
“R = Analog Regulator"(Equip.1 only)
Dose Time1: 999 sec
Reg. Ctrl. Mode1:(Pos/Neg): Pos
Setp. Init.Ch.1: 999,9 kg
Max. Alarm1: ????? ???
Min. Alarm1: ????? ???
Equipm.: 1 2 3 4Mode: R D 0 1
S
Vol. Ratio1: 1,000
“D = Dependant"(Equip. 2-3)
min. Volume1: 999,9 kg
Setpt. Start-Up1: 999,9 kg
Return to Config.Start = Enter
("???" Value and unit depending on selected sensor type)
Stroke Time1: 9,9 sec
Pause Time1: 9,9 sec
Equipment Mode1: (I/O): I
S
P = Permanent
Select Pump Type1: (S/P): S
Timed Pump Prod.1: 999,9 l/h
Stroke Pump Prod.1: 9999 p/l
Flowmeter Prod..1: 9999 p/l
only for timeddosing pump
only for stroked
dosing pump
only for product
flow meter
“F = Volume fix"(Equip. 1-3)
Equipment Number+1 (indexes to the next equipment number)
Prod. Flowmeter1: (Y/N):Y
Equipment 1-3only
Prod. Flow Rate=1: 9999,9 kg/h
Equip.1-3only
Equip.4
only
Operation Mode1:(R/C/F/D): R
Equip.1-3only
Equip.1-3 only
Max. Prod. Flow1: 999,9 kg/h
Min. Prod. Flow1: 999,9 kg/h
only whenflow meter = Yes
S E
Interval Time1: 999 sec
fixed Volume1: 999,9 kg
“C = Vol.fix + 2nd Crtl"(Equip. 1 only)
if CAN-Mode="Slave"
“1"
E
(Dosing Control with the Optical relays)
(T.I. = Teach In orEquipmentCalibration function)
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 58
Appendix D: Parameter List
Display Description Range Physicalunit
Alarm Delay Time Time Delay until a alarm is released if deviationbetween primary and secondary sensor exceedspreset values
0... 9,999 sec
Analog Input X Input current to analog input "X" (X=1,2) duringI/O-Test
0... 20 mA
ClO2 Scale max. ClO2-concentration necessary at sensor to give20mA output current (For values below 100ppmthe resolution will be 2 digits).
0,00 ...99,99100,0..200,0
ppm
Cond. Free curve1-x:____mS=__,_%
A custom specific characteristic to displayconductivity has been chosen. Every point "x" ofthis curve can be set independent of each other. Atotal of 7 points can be set. The conductivity hasto be set first.
0... 99990... 9999
µSmS
Cond. Free curve1-x:____mS=_,_%
A custom specific characteristic to displayconductivity has been chosen. The concentrationof product has to be set second.
0... 99,9 %
Cond. Ref. Temp. This is the reference temperature of theconductivity sensor (taken from its data sheet) ifthe sensor has no internal temperaturecompensation
0... 50 °C
Cond. Scale max. Conductivity (or concentration) necessary atsensor to give 20mA output current.
0... 99990... 99990... 99,9
µSmS%
Cond. Sel. Unit Conductivity can be displayed as "mS", "µS" or asa choice of different product concentrations (e.g.NaOH). A custom specific characteristic can beselected too ("free")
mS, NaOH,...,free
-
Cond. Temp.Coeff Temperature coefficient of conductivity sensor (ifwithout internal temperature
-9,99... +9,99 %/°C
Cond. Temp.Comp. This is a selection activates/ deactivates theadditional temperature compensation feature usinga temperature sensor at the second input
Y/ N -
Consumption/day One daily consumption counter for each product,these counters can be reset individually in mainmenu
0... 9.999,9 kg
Consumption/tot. One total consumption counter for each product,these counters can be together in configurationmenu
0... 99.999,9 kg
Deadband Difference between stopping the equipment 1 andrestart of dosing.
0... 100 %
Deviation max. Maximum deviation between two sensors ifsecondary control has been activated. An alarm isreleased if the measured value of the secondsensor is higher than (first sensor + max.
0... 100 %
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 59
deviation)
Deviation min. Minimum deviation between two sensors ifsecondary control has been activated. An alarm isreleased if the measured value of the secondsensor is lower than (first sensor - min. deviation)
0... 100 %
Dig. Input 1-8 The actual state of up to 8 digital inputs isdisplayed. As D-Determinator uses only 5 inputs,the upper 3 inputs are denoted as "X". A high-Level input signal will be displayed as "1", a lowlevel input signal will be displayed as "0".
0/1 -
Direct Display If no physical value is related to the sensor signalthe value can be displayed as "0/4-20 mA" or as"0-99,9 %"
mA, % -
Enter weight The amount of product (weight not volume) thathas been pumped during Teach-In Mode.
0...999,999 kg
Equipm.: 1 2 3 4 Operating mode of all 4 equipment displayedtogether.R (Equip. 1 only): Analog regulatorF (Equip. 1-3): Volume fixD (Equip. 2-3): DependantP or V (Equip. 4): Pump or Valve0 (Equip. 1-4): Equipment is not in use
R, F, D, 0, 1 -
Equipment Mode This is the pump type of the selected equipment:S: strokedP: permanent
S/ P -
Fixed Volume This is the volume that has to be delivered duringone interval in "volume fix" mode
0... 999,999 kg
Flow Sensor 1-8 The actual state of up to 8 flow sensor inputs isdisplayed. As D-Determinator uses only 3 flowsensors, the upper 5 inputs are denoted as "X". Ahigh-level input signal will be displayed as "1", alow level input signal will be displayed as "0" anda pulsing signal at the flow meter input will bedisplayed as "P"
0/ P/ 1 -
Flowmeter Prod. The number of pulses per liter produced by theflow meter taken from flow meter data sheet
0... 9999 Imp./liter
Interval Time This is the interval length in "Volume fix" mode.In this mode the pump is delivering the set "FixedVolume" within every interval and then will bestopped for remaining time.
0... 9.999 sec
Manual Vol=___,_ The remaining volume in "manual" mode. This isa read-only value.
0... 999,999 kg
Manual Volume This is the volume that has to be delivered during"manual" mode. This value is entered whilemanual mode is started
0... 999,999 kg
Max. Alarm This is the alarm threshold for the controlledvalue. If the value stays above this level longerthan the set delay time, an alarm is released.
See sensor scalemax.
Same assetpoint
Max. Prod. Flow If a flow meter is used the product flow is 0... 999,9 kg/h
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 60
controlled to stay within limits. This is themaximum flow. Outside these limits an alarm isreleased.
Min. Alarm This is the alarm threshold for the controlledvalue. If the value stays below this level longerthan the set delay time, an alarm is released.
See sensor scalemax.
Same assetpoint
Min. Prod. Flow If a flow meter is used the product flow iscontrolled to stay within limits. This is theminimum flow. Outside these limits an alarm isreleased.
0... 999,9 kg/h
min. Volume This is the minimum volume in "dependant"mode. In this mode the pump is running for the set"Pulse Time" and then will be stopped for "PauseTime".
0... 999,999 kg
Min.-Max.-Alarm This is a selection activates/ deactivates theadditional minimum-maximum alarm feature
Y/ N -
Network ID This is the address of this D-Determinator withinthe CAN-Network
0... 99 -
Operation Mode This is the operating mode of the equipment:R. analog regulatorC: Volume Fix + 2nd ControlF: volume fixD: dependant
R/ C/ F/ D -
Pause Time Eq. X This is the minimum pause time between twostroke pulses if the equipment (pump) is in"stroked" mode.
0... 9.999 sec
Pause Time This is the pause in "analog regulator" mode. Inthis mode the pump shall not run permanently ifthe setpoint is not reached
0... 9.999 sec
pH Scale max. pH-Value necessary at sensor to give 20 mAoutput current.
0... 14,0 pH
pH Scale min. pH-Value necessary at sensor to give 4 mA or 0mA, depending on sensor minimum outputcurrent.
0... 14,0 pH
Prod. Flowmeter This is a selection whether a flow meter forproduct is used
Y/ N -
Prod. Flow Rate
X: T.I.(Y/N):This is a selection if a flow data is enteredmanually or if a teach-in has to be made.
Y/ N -
Product Density The density of the used product, this parameter isnot necessary if a teach-in is made.
0... 1,999 kg/l
Puls Time This is the pulse length in "analog regulator"mode. In this mode the pump is running for the set"Pulse Time" and then will be stopped for "PauseTime".
0... 9.999 sec
RedOx Scale max. necessary at sensor to give 20 mA output current. 0...14,0 pH
RedOx Scale min. The pH-Value necessary at sensor to give 0/4 mAoutput current.
0...14,0 pH
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 61
Sel. Current min. The minimum current produced by the sensor canbe chosen between 0 and 4 mA. In case of a 4-20mA-Sensor an automatic control for broken wireis activated.
0/4 mA
Select Equipment The equipment to be configured can be selected. 1.. 4 -
Select Pump Type The type of pump used can be selected between"stroke" and "permanent"
S/ P -
Set Output 1-8 The actual state of 8 digital outputs is displayedand set. A high-Level output signal will bedisplayed as "1", a low level output signal will bedisplayed as "0".
0/1 -
Set Output 9-16 The actual state of up to 8 digital outputs isdisplayed and set. As D-Determinator uses only8+2 inputs, the upper 6 inputs are denoted as "X".A high-Level output signal will be displayed as"1", a low level output signal will be displayed as"0".
0/1 -
Setpoint This is the setpoint for the controlled value. If thevalue does not reach this level within the settimeout delay, an alarm is released.
See sensor scalemax.
Same assetpoint
Setpt. Init. Ch. This is the volume to be delivered during initialcharge mode. If the value reaches the alarm withinthis limit, an alarm is released.
0... 999,999 kg
Setpt. Start-Up This is the volume to be delivered during initialcharge mode. If the value reaches the alarm withinthis limit, an alarm is released.
0... 999,999 kg
Stroke Pump Prod The number of pulses necessary to pump one literof product by the dosing pump taken from dosingpump data sheet
0... 9999 Imp./liter
Stroke Time Eq.X This is the stroke pulse length if the equipment(pump) is in stroked mode.
0... 9.999 sec
Temp. Scale max. Temperature necessary at sensor to give 20 mAoutput current. This value is set in steps of 10°C.
0... 150 °C
Timed Pump Prod. The flow of product by the dosing pump at fullspeed taken from dosing pump data sheet
0... 999,9 Liter/h
Timeout Setpt. If the setpoint is not reached within this time, awarning is released.
0... 9.999 sec
Volume Ratio The ratio between product delivered by equipment1 and equipment 2/3 in "dependant" mode
0... 1,000 -
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 62
Appendix E: Mounting Diagram
236.70
203.35
123.35 23.35
323
300
(All Dimensions in Millimeters)
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 63
Appendix F: Specific conductivity diagrams
Conductivity for different solutions
Concentration. % NaOH % KOH % H3PO4 % H2SO4 % HNO30% 0,0 mS/cm 0,0 mS/cm 0,0 mS/cm 0,0 mS/cm 0,0 mS/cm1% 45,0 mS/cm 33,8 mS/cm 9,4 mS/cm 40,6 mS/cm 54,0 mS/cm2% 89,0 mS/cm 63,6 mS/cm 15,2 mS/cm 71,8 mS/cm 108,0 mS/cm3% 125,0 mS/cm 83,4 mS/cm 20,3 mS/cm 98,4 mS/cm 158,4 mS/cm4% 157,0 mS/cm 101,3 mS/cm 24,2 mS/cm 130,0 mS/cm 212,4 mS/cm
0
50
100
150
200
0% 1% 2% 3% 4% 5%
Concentration (%)
Co
nd
icti
vity
(m
S/c
m)
% NaOH
% KOH
% H3PO4
% H2SO4
% HNO3
Data taken at 20°C, measuring frequency 40 Hz
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 64
Relationship between RedOx potential and ClO2-Concentration
680 mV
700 mV
720 mV
740 mV
760 mV
780 mV
800 mV
820 mV
0.00 ppm 0.50 ppm 1.00 ppm 1.50 ppm 2.00 ppm 2.50 ppm
ClO2
ppm ClO2 RedOx-Pot.0,06 ppm 700 mV0,09 ppm 721 mV0,18 ppm 742 mV0,43 ppm 763 mV0,82 ppm 778 mV1,06 ppm 785 mV1,58 ppm 794 mV2,21 ppm 801 mV
D-Determinator (Software Version 1.21) 12/2002 FB5110INTissBPage 65
Glossary